
C-1
C Alarm Messages and Prompt Information
C.1 Physiological Alarm Messages
Note: XX represents a parameter label, such as PR, SpO
2
, NIBP, etc.The “L” field indicates
the alarm level. H means high, M means medium and L means low.
Alarm message L Cause Action
XX TOO HIGH M XX value exceeds the upper alarm
limit.
XX TOO LOW M XX value is lower than the lower
alarm limit..
Make sure the alarm
limits are appropriate for
the patient, and check
the patient’s condition.
NO PULSE H The pulse signal of the patient is so
weak that the monitor cannot
perform pulse analysis.
Check the connection of
the sensor and the
patient’s condition.
C.2 Technical Alarm Messages
Note: XX represents a parameter module such as NIBP or SpO
2
,
or a parameter label, such
as PR and SpO
2
. ## stands for patient category, i.e. adult, pediatric or neonate. The “A”
field indicates whether all alarm indications can be cleared or not, and the “B” field
indicates whether all alarm indications except the alarm message can be cleared or not. The
“L” field indicates the alarm level: H means high, M means medium and L means low. “*”
means the alarm level is user-adjustable.
Commentaires sur ces manuels