3.2.1 Inspection
Make sure the qualified service personnel have implemented a complete inspection before
putting the monitor into operation, after monitor servicing or system upgrading, or after the
monitor has been used for 6-12 consecutive months. This is to ensure the normal operation
of the system.
Check that
The environment and the power supply meet the specified requirements.
The monitor surface is free from stains.
The monitor surface, keys, connectors and accessories have no signs of physical
damage.
The power cords are not worn and the insulation is in good performance.
The grounding cables are properly connected.
Only specified accessories are applied.
The monitor clock is correct.
The audible and visual alarms are normal.
The recorder functions normally and the recorder paper meets the requirement.
The monitoring functions of the system are in good performance.
In case of any damage or exception, do not use the monitor. Contact the biomedical
engineers of the hospital or our Customer Service immediately.
3.2.2 Cleaning
WARNING
Be sure to shut down the system and disconnect all power cords from the outlet
before cleaning the equipment.
Your equipment should be cleaned regularly. If there is heavy pollution or lots of dust and
sand in your place, the equipment should be cleaned more frequently. Before cleaning the
equipment, consult your hospital’s regulations for cleaning, disinfecting and sterilizing
equipment.