9-3
9.3.2 Oral Temperature Measurement
Follow the procedure below to measure oral Temperature.
1. Make sure that the oral/axillary probe is connected to the probe connector and the
indication lamp beside the temperature unit lights to indicate that TEMP module
properly operates.
2. Select desired temperature type (the following procedure taking predict mode as an
example) and set TEMP POSITION to ORAL.
3. Unplug the probe from the probe sheath and insert it into a cover in the probe cover
box. Press the probe handle down firmly until the cover engages with the probe.
4. After the probe warmingup is ready, apply it under the patient’s tongue from either
side of the mouth. Verify that the probe reaches the rear sublingual pocket. Have the
patient close his/her lips to hold the probe.
5. Hold the probe in place. Make sure that the probe contacts with the patient’s oral tissue
throughout the measurement.
6. The monitor will give a beep as the temperature measurement is complete. The
temperature reading displays continuously.
7. Withdraw the probe from the patient’s mouth when accurate temperature reading is
obtained. Press firmly the ejection button on the top of the probe to eject the probe
cover. Return the probe into the sheath.
Temperature reading displays when final temperature is reached.
9.3.3 Axillary Temperature Measurement
1. Make sure that the oral/axillary probe is connected to the probe connector and the
indication lamp beside the temperature unit lights to indicate that TEMP module
properly operates.
2. Select desired temperature type (the following procedure taking predict mode as an
example) and set TEMP POSITION to AXILLARY.
3. Unplug the probe from the probe sheath and insert it into a cover in the probe cover
box. Press the probe handle down firmly until the cover engages with the probe.
4. After probe warmingup is ready, lift the patient’s arm to show the entire armpit. Apply
the probe as high as possible in the armpit. Check that the probe tip is completely
surrounded by the axillary tissue. Lower the patient’s arm so that it is tightly placed at
the patient side. Keep the patient’s arm and the probe in place throughout the
measurement.