3.2.3 Disinfection
WARNING
Disinfection may cause damage to the equipment; therefore, when preparing to
disinfect the equipment, consult your hospital’s infection controllers or
professionals.
The cleaning solutions above can only be used for general cleaning. If you use
them to control infections, we shall assume no responsibility for the
effectiveness.
Disinfection may cause damage to the equipment. We recommend the sterilization and
disinfection are contained in the hospital’s servicing schedule only when necessary. The
equipment should be cleaned prior to sterilization and disinfection.
Recommended sterilization material: Alcohol based (Ethanol 70%, Isopropanol 70%), and
aldehyde based.
CAUTION
ALWAYS dilute the solutions according to the manufacturer’s suggestions and
adopt lower concentration if possible.
NEVER submerge the equipment into water or any solution, or pour water or
any solution on the equipment.
ALWAYS wipe off all the excess liquids on the equipment surface and accessory
surface with a dry cloth.
NEVER use EtO and undiluted formaldehyde to disinfect the equipment.
Never permit high-pressure and high-temperature disinfection of the equipment
and accessories.