A-6
A.6.3 Nellcor SpO
2
Specification
Item Specification
Sensor Range Accuracy*
MAX-A, MAX-AL, MAX-N,
MAX-P, MAX-I, MAX-FAST
70% to 100%
0% to 69%
±2%
Undefined
OxiCliq A, OxiCliq N, OxiCliq
P, OxiCliq I
70 %to 100%
0% to 69%
±2.5%
Undefined
D-YS, DS-100A, OXI-A/N,
OXI-P/I
70 %to 100%
0% to 69%
±3%
Undefined
SpO
2
measurement range
and accuracy
MAX-R, D-YSE, D-YSPD
70% to 100%
0% to 69%
±3.5%
Undefined
PR measurement range and
accuracy
20 to 250 bpm: ±3 bpm
251 to 300 bpm: Undefined
Update period 1s
*: When sensors are used on neonatal patients as recommended, the specified accuracy range is
increased by ±1%, to account for the theoretical effect on oximeter measurements of fetal
hemoglobin in neonatal blood.
A.7 NIBP Specification
Item Specification
NIBP
Applicable standards IEC 60601-2-30, EN 1060-1, EN1060-3, AAMI SP-10
Method Oscillometry
Displayed parameters
Systolic pressure, diastolic pressure and mean pressure, pulse
rate
Mode of operation Manual, auto and STAT
Adult Pediatric Neonate
Systolic
pressure
40 to 270
mmHg
40 to 200
mmHg
40 to 135
mmHg
Diastolic
pressure
10 to 210
mmHg
10 to 150
mmHg
10 to 100
mmHg
Measurement range in
normal mode
Mean
pressure
20 to 230
mmHg
20 to 165
mmHg
20 to 110
mmHg
Accuracy
Maximum average error: ±5mmHg
Maximum standard deviation: 8mmHg