9-2
9.2 Temperature Setup
In the SYSTEM SETUP menu, select TEMP SETUP, you can adjust:
TEMP TYPE: PREDICT or MONITOR. If MONTIOR is selected, The LED
indicating measuring mode on the front panel illuminates.
TEMP POSITION:: ORAL, AXILLARY or RECTAL. Corresponding indicating lamp
on the front panel illuminates if a measurement site is selected.
TEMP UNIT: or .
9.3 Monitoring Procedure
9.3.1 TEMP position
The TEMP module can be configured with 2 types of TEMP probe: oral/axillary probe (blue)
and rectal probe (red). The blue oral/axillary probe shall only be used with blue probe
sheath, while the red rectal probe shall only be used with red sheath. Be sure to select
correct probe.
Oral/axillary probe: This type of probe is intended for taking oral or axillary
temperature of adult and pediatric patients, or axillary temperature of neonatal patient.
Rectal probe: This type of probe is intended for taking rectal temperature of adult and
pediatric patient.
NOTE
If the patient type is adult or pediatric, the equipment automatically selects oral
as the measurement site when the oral/axillary temperature probe is in use. You
can change the measurement site in the TEMP SETUP dialog.
If the patient type is neonatal, the equipment automatically selects axillary as
the measurement site if the oral/axillary temperature probe is in use. In this
case, you cannot change measurement site.