C-9
Alarm message A B L Cause Action
TEMP WRONG
PROBE
No No H A TEMP probe not
supplied by our
company is used.
Replace with a
TEMP probe we
supply.
TEMP COMM
ERR
No No H TEMP module is not
available or TEMP
module fails
Check if a TEMP
module is
available. If yes,
replace the TEMP
module.
C.2.7 Recorder Module Alarm Messages
Alarm message A B L Cause Action
RECORDER INIT
ERR N
Yes No M An error occurs
during the recorder
initialization.
Contact the hospital’s
engineers or our
Customer Service.
Note: N represents the error number.
REC SELFTEST
ERR
Yes No M An error might
occur to the RAM,
ROM and CPU
watchdog.
Restart the recorder.
If the error remains,
contact our service
personnel for repair.
RECORDER VLT
HIGH
No No H
RECORDER VLT
LOW
No No H
A problem occurs to
the system’s power
supply.
If this alarm message
occurs frequently,
contact our service
personnel for repair.
RECORDER
HEAD HOT
No No H The thermal head of
the recorder is too
hot.
Resume the recording
after the recorder
cools down
completely. If the
problem still exists,
contact our service
personnel for repair.
REC HEAD
WRONG POS.
Yes Yes L The thermal head of
the recorder is in
wrong position.
Restore the control
lever of the recorder
to its previous
position.
REC OUT OF
PAPER
Yes Yes L The recorder paper
is used up.
Replace with a new
paper roll.