5.3 Alarm Status
Apart from the aforementioned alarm indicators, the patient monitor still uses symbols to
indicate the alarm status: You can set alarms to the following status as desired:
Audio off:
is displayed at the right side of patient information area and a
prompt message “AUDIO ALARM OFF” appears at the right side of the alarm
information area.
Physiological Alarm off:
is displayed at the right side of patient information area
and a prompt message “ALARM OFF” appears at the right side of the alarm
information area.
Alarm paused:
is displayed at the right side of patient information area ; a
prompt message “ALARM PAUSE” and remaining alarm pause time appear at the
right side of the alarm information area.
Alarm silenced:
is displayed at the right side of patient information area.
5.3.1 Audio Off
To switch off alarm tones, follow this procedure:
1. Select SYSTEM SETUPMAINTAINUSER MAINTAIN; enter required password
and set MIN ALARM VOL to 0 in the pop-up menu.
2. Select SYSTEM SETUPCOMMON SETUP; set ALMAR VOL to 0 in the pop-up
menu. Thus alarm tones are switched off.
If AUDIO OFF PROMPT in the USER MAINTAIN menu is set to ON, the monitor clanks
at an interval of one minute. The volume level of prompt tone is 2.
The monitor automatically exit the audio off status when:
ALARM VOL is manually changed to a level other than 0;
The monitor is restarted after being turned off for more than 120s, or when patient ID
or patient type is changed in the case it is preset to load factory configuration or user
configuration at startup.
LOAD FACTORY CONFIG or LOAD USER CONFIG is selected in the DEFAULT
menu.