5-5
WARNING
If alarm tones are switched off, the monitor does not give audible alarm signal
even if new alarms occur. So take care when you switch off alarm tones.
5.3.2 Physiological Alarm Off
If alarms related to a parameter are switched off, the monitor does not generate alarms even
if the measured parameter values exceed the alarm limit. This status is called Alarms Off.
To disable the alarms of a parameter, you need to open SET ALARMS menu .Take NIBP
(Non-Invasive Blood Pressure) as an example.
1. Press
to open the SET ALARMS menu.
2. Move the cursor to the pane to the right of SYS ALM.
3. Select OFF and the NIBP alarms are disabled.
NOTE
Within 30s after the NIBP, SPO2 and TEMP modules are loaded, physiological
and technical alarms related to these modules are switched off automatically.
5.3.3 Alarms Paused
Shortly pressing for less than 2 seconds can pause all alarms for 2 minutes. When
alarms are paused,
Alarm lamp flashing and alarm tones are suspended.
For a physiological alarm, alarm message is not displayed.
For a technical alarm, alarm message is displayed if Alarm lamp flashing and alarm
tones can be cleared.
For a technical alarm, alarm message is not displayed if all alarm indications can be
cleared completely.
The remaining alarm pause time is shown at the right side fo alarm information area.