5-6
In the alarm paused status,
All alarm indications are disabled when a new physiological alarm occurs.
Alarm lamp flashing and alarm tone are disabled and only alarm message is displayed
when a new technical alarm occurs.
The monitor automatically exits the alarm paused status when the alarm pause time expires.
You can also press
to manually leave the larm paused status.
5.3.4 Alarm Silenced
You can silence alarms by press and hold for more than 2 seconds. In the alarm
silenced status, alarm indications except alarm tones operate properly. The monitor
automatically exits the alarm silenced status if a new technical or physiological alarm
occurs.
5.3.5 Status Switchover
1. In the normal status,
Press
for less than 2 seconds to switch the monitor to the Alarms Paused
status, or
Press
for 2 seconds or more to switch the monitor to the Alarms silenced
status.
2. In the alarm paused status,
Press
for less than 2 seconds to switch the monitor to the normal status, or
Press
for 2 seconds or more to switch the monitor to the Alarms silenced
status.
3. In the alarm silenced status,
Press
for less than 2 seconds to switch the monitor to the Alarms Paused
status, or
Press
for 2 seconds or more to switch the monitor to the normal status.