Mindray DPM 1 Manuel de l'opérateur Page 46

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 120
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 45
4-6
WARNING
Then nurse call settings shall not be changed by non-medical staff.
4.2.6 Version
Select SYSTEM SETUPMAINTAINVERSION, you can view information on the
monitors hardware version and software version.
4.2.7 Configuration
4.2.7.1 Presetting Configuration
If the monitor is turned off for less than 60 seconds, last configuration will be loaded
automatically when it is restarted. If it has been turned off for more than 120 seconds,
different configuration will be loaded according to the setting when it is restarted, or when
patient type or patient ID is changed.
1. Select SYSTEM SETUPMAINTAINUSER MAINTAIN; enter required password
to enter the USER MAINTAIN menu.
2. Select SELECT CONFIG.
3. Select LAST CONFIG, USER CONFIG or FACTORY CONFIG as desired.
4. Select OK.
NOTE
When the monitor is restarted after being turned off for 60 to 120 seconds, the
configuration loaded may be last configuration, user configuration or factory
configuration
4.2.7.2 Saving User Configuration
1. Make sure the modification you make is appropriate and correct.
2. Select SYSTEM SETUPMAINTAINUSER MAINTAIN; enter required password
to enter the USER MAINTAIN menu.
3. Select SAVE USER CONFIG.
Vue de la page 45
1 2 ... 41 42 43 44 45 46 47 48 49 50 51 ... 119 120

Commentaires sur ces manuels

Pas de commentaire