4-9
DELETE: By selecting DELETE, you can toggle between CURRENT ITEM,
ITEMS OF CURRENT ID, or ITEMS OF ALL ID. Then select
OK to respectively delete the current trend data, all trend data of
current ID, or data of all ID.
4.5 Standby State
4.5.1 Entering the Standby State
To enter the Standby state, press for less than 2 seconds and then select OK in the
CONFIRM STANDBY STATE dialog box.
4.5.2 Exiting the Standby State
In the Standby state, press any key on the front panel of the monitor. The EXIT STANDBY
dialog box appears. Select YES to exit the Standby state. If no operation is done within 30
seconds, the monitor will automatically select NO, this dialog box will disappear, and the
monitor will keep in the Standby state.
In the following circumstance,
The monitor presents a prompt to ask if you want to exit the Standby mode when a
button is pressed.
The monitor exits the Standby mode when a bar code scanner is used to identify
patient ID.
The monitor exits the Standby mode automatically after continuously receiving SpO
2
signal for 5 seconds.
The monitor exits the Standby mode automatically when the TEMP probe is
withdrawn from the probe sheath.
In the case that the monitor runs on internal battery, the monitor exits the Standby
mode and presents an alarm “BAT. VOLTAGE LOW” when the battery power is low.