3-6
3.1.5.5 Connecting the network cable
The network connector of the monitor is a standard RJ45 connector. It connects the monitor
with the specified central monitoring system, or with a PC for software upgrade and data
export. To connect the monitor with CMS or PC,
1. Connect one end of the network cable to the network connector of the monitor.
2. Connect the other end of the network cable to the hub of the central monitoring system,
or to the network connector of a PC.
WARNING
Upgrading the system through network is only to be executed by personnel we
authorize.
3.1.5.6 Nurse call connector
The nurse call connector is used for the nurse call function. If connected to the hospital’s
nurse call system through a special nurse call cable, the monitor can generate nurse call
signals when an alarm occurs. The output end of the nurse call cable consists of two free
cords that is neutral. The nurse call system shall be installed by the maintenance engineer
from the manufacturer or the engineer of the hospital.
3.1.5.7 Connecting the Bar Code Scanner
The bar code scanner shall be connected to the RS232 connector. Use only bar code
scanners supplied by us.