C-11
C.2.8 System Alarm Messages
Alarm message A B L Cause Action
REAL CLOCK
NEED SET
No No H The system time is
incorrect.
Reset the system time
and then restart the
monitor.
REAL CLK NOT
EXIST
No No H There is no button
battery, or the
battery power is
depleted.
Install a button
battery or replace
with a new one.
NET INIT ERR (G.)
No No M
NET INIT ERR
(Ram)
No No M
NET INIT ERR
(Reg)
No No M
NET ERR (Run 1) No No M
NET ERR (Run 2) No No M
The monitor cannot
be connected to the
network due to
network problem.
Contact our service
personnel for repair.
12V TOO HIGH No No H
12V TOO LOW No No H
A problem occurs to
the system’s power
supply.
If this alarm message
occurs frequently,
contact our service
personnel for repair.
BATTERY TOO
LOW
No No H The battery voltage
is too low.
Connect the monitor
to AC mains to
charge the battery.
DATAOUT FAIL Yes Yes M A failure occurs in
data transmission
during the data
output.
Re-output the data.
NET ERROR Yes Yes M The monitor is not
connected to the
network
Check the network
connection.
BARCODE ID
FAILED
Yes No L Bar code format
cannot be decoded.
Enter patient ID
manually or use a
decodable bar code.