4.2.5 Nurse Call Setup
Select SYSTEM SETUPMAINTAINNURSE CALL. You can set the following items
in the pop-up menu:
SIGNAL DURATION
1. CONTINUUM
It indicates that the nurse call signal duration is the same with the alarm duration,
namely, the nurse call signal lasts from the beginning of the alarm to the end of the
alarm.
2. PULSE
It indicates that the nurse call signal is a pulse signal whose duration is 1s. When
multiple alarms occur, the monitor outputs only one pulse signal; if another alarm
occurs before the current alarm is cleared, the monitor will output another pulse signal.
SIGNAL TYPE
1. NORMAL CLOSE: set the signal type to NORMAL CLOSE when the nurse call
system of the hospital is set to normally-closed;
2. NORMAL OPEN: set the signal type to NORMAL OPEN when the nurse call system
of the hospital is set to normally-open.
NORMAL CLOSE and NORMAL OPEN are technical terms which describe the type of
relay in nurse call system. The hospitals should select the right signal type consistent with
their nurse call system.
ALM LEV: HIGH, MED, LOW; check box
ALM TYPE: TECH, PHYS; check box.
The system will send the nurse call signal according to the selected alarm level and alarm
type. If neither alarm level nor alarm type is selected, the system will not send any nurse
call signal when alarms occur.
The Nurse Call function doesn't distinguish among ALM LEV and ALM TYPE. As long as
any type is checked and when the alarmable event occurs, the monitor will send the same
signal to the interface of the Nurse Call System.
NOTE
The medical/nursing staff are not expected to take the nurse call function as the
major alarm notification. The patient conditions should be determined based on
the audible/visual alarm of the monitor and the clinical symptoms of the patient.
In the Alarm Paused or Standby State, the nurse call function will be disabled.