3-9
The exterior surfaces of the equipment may be cleaned with a clean and soft cloth, sponge
or cotton ball, dampened with a non-erosive cleaning solution. Drying off excess cleaning
solution before cleaning the equipment is recommended. Following are examples of
cleaning solutions:
Diluted soap water
Diluted ammonia water
Diluted sodium hypochlorite (bleaching agent)
Diluted formaldehyde (35 to 37%)
Hydrogen peroxide (3%)
Ethanol (70%), or Isopropanol (70%)
NOTE
The sodium hypochlorite of the concentration ranging from 500ppm (1:100
diluted bleaching agent for home use) to 5000ppm (1:10 diluted bleaching agent
for home use) is effective. The required concentration depends on the quantity
of the organic substances (such as blood, mucus) on the equipment surface.
To avoid damage to the equipment, please follow these rules:
ALWAYS dilute the solutions according to the manufacturer’s suggestions.
ALWAYS wipe off all the excess cleaning solution with a dry cloth after cleaning.
NEVER submerge the equipment into water or any cleaning solution, or pour or spray
water or any cleaning solution on the equipment.
NEVER permit fluids run into the casing, switches, connectors, or any ventilation
openings in the equipment.
NEVER use abrasive or erosive cleaners of any kind as well as cleaners containing
acetone.
CAUTION
Failure to follow these rules may erode or fray the casing, or blur lettering on
the labels, or cause equipment failures.
For cleaning information of accessories, please refer to the chapters for specific patient
parameters and the instructions for use of the accessories.