C-4
C.2.3 Mindray DS SpO
2
Module Alarm Messages
Alarm message A B L Cause Action
SPO2 SENSOR
OFF
No Yes * The sensor is
disconnected from
the patient or the
monitor.
Make sure that the
sensor is placed at an
appropriate position
and the monitor is
connected to cables
correctly.
SPO2 LOW
PERFUSION
No No L The pulse signal is
too weak.
Move the sensor to a
site with better
perfusion.
SpO
2
sensor is
detached from the
patient or monitor,
or SpO
2
sensor is
not properly
connected.
Disconnect SpO
2
sensor and reconnect
it as per instructions
for use. If the alarm
persists, SpO
2
sensor
or cable may be
damaged.
SPO2 NO SENSOR
No Yes L
SpO
2
sensor is
reversely connected
Disconnect SpO
2
sensor and reconnect
it as per instructions
for use. Pay attention
to the mark on the
sensor.