8-7
8.4.4 NIBP Leakage Test
You can test the pump for air leakage by selecting SYSTEM SETUPMAINTAINNIBP
TOOLSNIBP LEAK TEST. With the NIBP cuff connected, you can test the air way for
leakage by selecting NIBP LEAK TEST. If the NIBP leakage test is passed, no prompt will
be given; If it fails, error information will be displayed in the NIBP parameter area.
To test air leakage, follow the procedure below:
1. Set the patient category to adult.
2. Connect a rigid metal container or vessel with a capacity of 500 ml ± 5% to the NIBP
cuff connector of the monitor.
Figure 8-2 NIBP Leakage Test
3. Select NIBP LEAK TEST in the M NIBP TOOLS menu, the message “Pneum
testing...” displays in the information area..
4. The cuff automatically deflates in 20s which means NIBP leakage test is completed.
5. If no message appears in the NIBP parameter area, it indicates the airway is in good
condition and no air leak exists. However, if the system prompts “PNEUMATIC
LEAK”, it indicates the airway may have an air leak. In this case, check for loose
connections. After confirming that all connections are secure, perform the test again.
If the problem persists, contact our Customer Service.
NOTE
z The leakage test is intended for use to simply determine whether there are
leakages in the NIBP airway. It is not the same as that specified in the EN 1060-3
standard.
Monitor
Connector for
NIBP
cu
f
f
Air tubing
Metal
vessel