C-7
C.2.5 Nellcor SpO
2
Module Alarm Messages
Alarm message A B L Cause Action
SPO2 SENSOR
OFF
No Yes * The sensor is
disconnected from
the patient or the
monitor.
Make sure that the
sensor is placed at an
appropriate position
and the monitor is
connected to cables
correctly.
The sensor is
disconnected from
the patient or the
monitor, or the
sensor is not
connected properly.
Disconnect and
reconnect the sensor
as directed by the
instructions. If the
alarm remains, the
sensor or the cable
might have been
damaged.
SPO2 NO SENSOR
No Yes L
SpO
2
sensor is
reversed.
Disconnect and
reconnect the sensor
as directed by the
instructions. Pay
attention to the mark
on the sensor.
SPO2
INTERFERENCE
No No L The pulse signals
are subject to great
external
interference.
Reduce or remove
external interference.
SPO2 BOARD
FAULT
No No H The SpO
2
board
malfunctions and
might be unable to
measure the pulse
signals correctly.
Stop using the SpO
2
module, and contact
biomedical engineers
or us for
maintenance.
SPO2 SENSOR
FAULT
No No H The sensor is
damaged.
Stop using the sensor.
SPO2 WEAK
SIGNAL
No No L The SpO
2
signal is
weak.
SPO2 WEAK
PULSE
No No L The detected pulse
signal is too weak.
Change the sensor
site for better signals.