8-6
8.4.3 Calibrating NIBP
In SYSTEM SETUP menu, select MAINTAINNIBP TOOLSNIBP CALIBRATE to
start NIBP calibration. Then the button NIBP CALIBRATE changes to STOP CALIBRATE.
By selecting STOP CALIBRATE, you can stop the calibration.
Follow this procedure to calibrate the cuff pressure with a calibrated reference manometer
(or mercury manometer) with accuracy 1mmHg:
1. Disconnect the NIBP cuff from the monitor and replace it with a metal vessel with
capacity 500±25ml.
2. Connect a calibrated reference manometer (with an error less than 0.8 mmHg), a ball
pump and a hoses to the monitor, as shown in the figure below:
3. Select NIBP CALIBRATE.
4. Inflate the metal vessel with the ball pump until the reference manometer reads 0, then
50, and finally 200 mmHg.
5. The difference between the indicated pressure of the reference manometer and that of
the monitor shall not exceed 3 mmHg. Otherwise, contact Our Customer Service.
Figure 8-1 NIBP Calibration
NOTE
z NIBP calibration shall be performed every two years or according to your
hospital’s protocol.
Metal vessel