A Product Specifications
A.1 Safety Specifications
Item Specification
Type of protection against
electrical shock
Class I; internally/externally powered equipment
When integrity of the external protection grounding or grounding
cable is doubtable, the equipment must be powered by the
internal power (battery).
Degree of protection
against electric shock
SpO
2
/NIBP/TEMP: CF
Degree of protection
against hazards of
explosion
Not suitable (ordinary)
Degree of protection
against harmful ingress of
water
Not suitable (ordinary)
Mode of operation Continuous
Equipment type Portable
A.2 Environmental Specifications
Item Specification
Operating ambient
temperature
0 to 40 ºC
10 to 40 ºC (50 to 104 ºF) (SmarTemp™ TEMP module)
Operating relative humidity 15 to 95%, non-condensing
Atmospheric pressure at
working condition
(operating altitude)
57107.4KPa
70 to 106 kPa(SmarTemp™ TEMP module)
Storage temperature -20 to 60 ºC
Storage relative humidity 10 to 95%, non-condensing
atmospheric pressure in
Storage (Storage altitude)
16107.4KPa
50106KPa(SmarTemp™ TEMP module)