4-8
4.4 Trend Screen
The Trend screen displays systolic pressure (S), diastolic pressure (D) and mean pressure
(M), SpO
2
, PR and TEMP, as shown in the figure below. Up to 1200 groups of data can be
stored.
The monitor stores trend data in either of the following modes:
For SpO
2
and TEMP values in monitor mode, the trend data is sampled value. The time
interval of sampling is 30 seconds. Additionally, the last value obtained before the
“SENSOR OFF” or “TEMP SENSOR OFF” alarm occurs will also be stored.
For NIBP and TEMP values obtained in predict mode, the trend data is the actual value
measured when the data is sampled.
NOTE
If the difference between the time when two storage modes happen is less than
one second, the trends displayed may be of the same time, but SpO2 trends
and/or TEMP trends in monitor mode may be different.
Figure 4-4
Press
to enter the REVIEW SETUP menu where you can change:
REVIEW ID: Select the patient ID you want to review.
REVIEW MODE: You can set REVIEW MODE to ALL, NIBP EVENTS or TEMP
EVENTS. If you select ALL, all the trend data of the selected
patient will be displayed. If you select NIBP EVENTS, the trend
data which includes all NIBP measurements of the selected patient
will be displayed. If you select TEMP EVENTS, the trend data
which includes all TEMP measurements of the selected patient
will be displayed.