C.3 Prompt Messages
Prompt message Cause Action
SEARCH PULSE The SpO
2
module is searching the
pulse.
Wait till the end of the search.
SPO2 MOTION SpO
2
sensor is moving. Eliminate sensor movement.
SPO2 Alarm Disabled SpO
2
alarm is switched off. Switch on SpO
2
alarm.
RECORDER BUSY The recorder is recording data. Wait till the end of the recording.
Manual measure... The NIBP module is performing a
manual measurement.
Cont measuring... The NIBP module is performing a
continuous measurement.
Auto measuring... The NIBP module is performing
an automatic measurement.
Wait till the end of the
measurement.
NIBP Alarm Disabled NIBP alarm is switched off. Switch on NIBP alarm.
Resetting… Module is resetting. Wait till resetting is finished.
Please start The interval for automatic
measurement has been selected.
Press to start the
measurement.
Calibrating... The NIBP module is being
calibrated.
Wait till the end of the
calibration.
Calibration over The calibration is finished. None
Testing leak... The NIBP module is testing for
air leakage.
Wait till the end of test.
Pneum test over The test for air leakage is
finished.
None
Measurement over
is pressed during
measurement to stop NIBP
measurement.
None
Reset failed
The NIBP module fails to be
reset.
None
TEMP Warming Up
TEMP module is warming up. Wait till TEMP module
completes warmup.
TEMP Prediction
Ready
TEMP module completes
warming up and predictive
measurement can be performed
now.
Perform predictive TEMP
measurement.