Mindray DPM 1 Manuel de l'opérateur Page 33

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 120
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 32
3-3
3.1.5 Installation Method
WARNING
Equipments connected to this monitor must be certified according to the
respective IEC standards (e.g. IEC 60950 for information technology equipment
and IEC 60601-1 for medical electrical equipment). Furthermore all
configurations shall comply with the valid version of IEC 60601-1-1. Any person
who connects additional equipment to the signal input or signal output is
responsible to ensure the system complies with the requirements of the valid
version of IEC 60601-1-1. If in doubt, contact our company or customer service.
If the monitor is connected to another electrical instrument and the instrument
specifications cannot tell whether the instrument combination is hazardous (e.g.
due to summation of leakage currents), you should consult us or experts in the
field to ensure the required safety of all instruments concerned.
NOTE
The following operations are not all required. User-customized installation by
authorized personnel is provided.
3.1.5.1 Connecting to AC mains
1. Use the original 3-core power cable.
2. Connect the power cable to the AC mains input connector on the rear panel of the
monitor.
3. Connect the other end of the power cable to a compatible 3-prong hospital grade AC
power outlet.
The 3-prong power outlet must be grounded. In case of any doubt, contact related personnel
of the hospital.
WARNING
Do not use three-wire to two-wire adapter with this instrument.
To avoid unexpected power interruption, do no use power outlet with a
wall-mounted switch control.
Vue de la page 32
1 2 ... 28 29 30 31 32 33 34 35 36 37 38 ... 119 120

Commentaires sur ces manuels

Pas de commentaire