C.2.6 SmarTemp™ TEMP Module Alarm Messages
Alarm message A B L Cause Action
WARMUP TIMED
OUT
Yes Yes M TEMP probe initial
temperature is too high.
Cool the TEMP
probe before taking
measurement.
WARMING
RESISTOR ERR
No No M The warming resistor in
the TEMP probe fails.
Replace the TEMP
probe.
TEMP PROBE
MISPLACED
Yes Yes M TEMP probe is not
placed in the sheath or
the probe sheath is not
in place.
1. Check that the
probe sheath is in
place.
2. Replace the
TEMP probe in the
sheath properly.
ENV TEMP
OVERRANGE
No Yes M The ambient
temperature is beyond
the measuring range.
Take measurement
in proper working
condition.
TEMP VOLTAGE
ERR
No Yes M Supply voltage is too
high or too low.
Check the power
supply.
TEMP NO PROBE
No Yes M The TEMP probe is
disconnected from the
TEMP module.
Reconnect the
probe with the
TEMP module.
TEMP
PREDICTION
ERR
Yes Yes L Improper temperature
measurement
Take TEMP
measurement again
correctly.
TEMP SELFTEST
ERR
No No H An error occurs during
the TEMP module
initialization
Replace the TEMP
module
TEMP PROBE
OFF
No Yes L TEMP probe does not
contact with the patient.
Take measurement
again after the
probe warms up.
TEMP OVER
HIGH LIMIT
No No H The temperature
measured is too high or
measurement error
Lower the
measured
temperature or
replace the TEMP
module.
TEMP OVER
LOW LIMIT
No No H The temperature
measured is too low or
measurement error
Raise the measured
temperature or
replace the TEMP
module.