4-21
1. ANALOG OUT (analog output)
If this option is selected, the auxiliary output port will be able to output analog signals,
and you can perform the settings in the ANALOG menu. For details, see
4.4.6 Analog
Output.
If this option is not selected, the analog output function will be disabled and the
ANALOG >> option in the SYSTEM SETUP menu will become inactive. In this
situation, you cannot set the information in the ANALOG menu.
2. NURSE CALL
If this option is selected, the auxiliary output port will be able to output nurse call
signals, and you can perform the settings in the NURSE CALL SETUP submenu of the
USER MAINTAIN menu. For details, see
4.7.4 Nurse Call Setup.
If not selected, the nurse call function will be disabled and the NURSE CALL
SETUP>> option in USER MAINTAIN will become inactive. In this situation, you
cannot set the information in the NURSE CALL SETUP submenu.
3. DEFIB. SYN (defibrillator synchronization signals)
If this option is selected, the auxiliary output port will be able to output defibrillator
synchronization signals. In this situation, you can turn on DEFIB SYNC in the ECG
SETUP menu to enable the defibrillator synchronization. For details, see
10.4 ECG
Setup Menu.
If not selected, the defibrillator synchronization function will be disabled and the
DEFIB SYNC option in the ECG SETUP menu will set to OFF (it will not be
user-adjustable).
LEAD NAMING Options: AHA and EURO;
POWERON CONFI
It indicates the configuration to be loaded when the monitor is
powered on. The options are LAST CONF, USER ADU
CONF, USER PED CONF and USER NEO CONF.
LAST CONF is the real-time configuration that is selected
before the monitor is last turn off. If the user configuration of
a patient type is not saved, the factory configuration of this
type will be loaded.
If the time interval between turnoff and turnon is less than 60
seconds, LAST CONF is automatically restored when the
monitor is restarted. If this time interval is greater than 120
seconds, the setting of POWERON CONF will be loaded
when the monitor is restarted.
SAVE AS USER
CONFIG
You can change the monitor’s settings and save the changed
settings as user configuration of this patient type so that you
can load the user configuration when a new patient is
admitted.
You can restore the saved user configuration in the CONFIG
menu or set it as POWERON CONFI.