A-10
Amplitude:
Width:
Rise time:
±4 to ±700 mV (3/5-lead)
±2 to ±700 mV (12-lead)
0.1 to 2 ms
10 to 100 µs
When tested in accordance with the ANSI/AAMI EC13: Sections
4.1.4.1 and 4.1.4.3, the heart rate meter rejects all pulses meeting
the following conditions.
Pulse rejection
Amplitude:
Width:
Rise time:
Min. input slew rate:
±2 to ±700 mV
0.1 to 2 ms
10 to 100 µs
20 V/s RTI
ST segment measurement
Measurement range -2.0 to +2.0 mV
Precision
-0.8 to +0.8 mV: ±0.02 mV or ±10%, whichever is greater
Beyond this range: Undefined
Update period Updated every 16 valid beats
A.9 RESP Specifications
Measurement technique Thoracic impedance
Lead Optional: lead I and lead II; default lead II
Respiration excitation
waveform
< 300 µA, sinusoid, 62.8 kHz (±10%)
Respiration impedance test
range
0.3 to 3 Ω
Baseline impedance range 200 to 2500 Ω (using an ECG cable with 1kΩ resistance)
Differential input
impedance
> 2.5 MΩ
Linear Signal Range 3 Ω p-p minimum
Bandwidth 0.2 to 2 Hz (-3 dB)
Sweep speed 6.25 mm/s, 12.5 mm/s, 25 mm/s
RR
Measurement range
Adult:
Pediatric/neonate:
0 to 120 BrPM
0 to 150 BrPM
Resolution 1 BrPM
Precision
7 to 150 BrPM:
0 to 6 BrPM:
±2 BrPM or ±2%, whichever is greater.
Undefined.
Apnea alarm delay 10 to 40 s