10-15
NOTE
z When monitoring a patient with a pacemaker, PACE must be turned ON.
Otherwise, the system will count the pacemaker pulse as QRS complex. Do not
completely depend on the alarms of heart rate. The patient with a pacemaker must
be nearly monitored.
z When PACE is turned ON, the system will neither detect the arrhythmia relating
to premature ventricular beats (including PVCs counting) nor perform the ST
analysis. When monitoring a patient without a pacemaker, PACE should be
turned OFF.
z The PACE tag can be printed in the real-time recording when PACE is turned
ON.
PACE LEAD Set PACE LEAD as required.
NOTE
z You can switch the leads by setting LEAD TYPE in the ECG SETUP menu. At the
time of switching between 5 LEADS and 3 LEADS, PACE LEAD in the ECG
SETUP menu will be II.
PACER RATE Some pace pulses may be difficult to suppress. When this
happens, the pace pulse may be counted as a QRS complex,
and could result in incorrect HR measurements and failure to
detect some arrhythmias. This option allows you to input the
heart rate of pacemaker, so that the software can calculate the
HR and detect the arrhythmias more accurately. When PACE
is set to OFF, this option is disabled.
NOTCH Determines whether filter or not.
ON: The monitor protects the signals from the noise
generated by the power line.
OFF: No filtering.