C-12
C.2.13 Recorder Module Alarm Messages
Alarm message A B L Cause Measure
RECORDER
INIT ERR N
Yes No 2 An error occurs during the
recorder initialization.
Contact the hospital’s
engineers or our customer
Service.
Note: N represents the error number.
RECORDER
SELFTEST ERR
Yes No 2 An error might occur to the
RAM, ROM and CPU
watchdog.
Open the RECORD menu and
select the CLEAR REC TASK
option. If the problem remains,
contact our company for repair.
RECORDER
VLT HIGH
No No 1
RECORDER
VLT LOW
No No 1
A problem occurs to the
system power.
If this alarm message is given
for many times, contact our
company for repair.
RECORDER
HEAD HOT
No No 1 The thermal head of the
recorder is too hot.
Resume the recording till the
recorder cools down
completely. If the problem still
exists, contact our company for
repair.
REC HEAD IN
WRONG POS.
Yes Yes 3 The thermal head of the
recorder is in wrong
position.
Restore the control lever of the
recorder to its previous
position.
RECORDER
OUT OF PAPER
Yes Yes 3 The recorder paper is used
up.
Replace with a new paper roll.
RECORDER
PAPER JAM
No No 2 The recording continues for
more than 30 minutes.
Place the recorder correctly and
try again.
RECORDER
COMM ERR
Yes No 2 Error in recorder
communication.
Open the RECORD menu and
select the CLEAR REC TASK
option. If the problem remains,
contact our company for repair.
TOO MANY
REC TASKS
No No 2 Quite a few alarm events
occur at the same time.
Check the patient’s condition
and the alarms. Open the
RECORD menu and select the
CLEAR REC TASK option. If
the problem remains, contact
our company for repair.
RECORDER
PAPER W.P.
Yes Yes 2 The paper roll of the
recorder is not placed in the
correct position.
Place the paper roll correctly.
RECORDER S.
COMM ERR
Yes No 2 Error in recorder
communication.
REC NOT
AVAILABLE
No No 2 Error in the recorder work
mode.
Open the RECORD menu and
select the CLEAR REC TASK
option. If the problem remains,
contact our company for repair.
Commentaires sur ces manuels