C-15
Prompt messages Cause Measure
AG IS STARTING The AG module is starting. Wait for the AG module to
finish the startup.
AG WARM UP The AG module is warming itself up. Wait for the AG module to
finish warming itself up.
NIBP module
Manual measure... The NIBP module is performing the
manual measurement.
CONTINUAL... The NIBP module is performing the
continuous measurement.
Auto measuring... The NIBP module is performing the auto
measurement.
Wait for the NIBP module to
finish the measurement.
Resetting...
Resetting...
The NIBP module is being reset. Wait the NIBP module to finish
the resetting.
Please start This message appears after the auto
measurement interval is selected.
Press the NIBP button to start
the measurement.
CALIBRATE... The NIBP module is performing the
calibration.
Wait for the NIBP module to
finish the calibration.
Calibration over The calibration is finished. None
PNEUMATIC... The NIBP module is checking the
pneumatic system for leakage.
Wait for the NIBP module to
finish checking the pneumatic
system.
Pneum test over The NIBP finishes checking the
pneumatic system for leakage.
None
Measurement over The NIBP button is pressed during the
measurement.
None
Reset failed The reset fails. None
ST LEARNING
ARR LEARNING
The QRS complex template for the ARR
analysis is forming.
Wait till the end of the ARR
learning.
DEFIB. SYNC.ON The DEFIB. SYNC. switch is turned on. It indicates that the auxiliary
output port is outputting
DEFIB. SYNC. signals.