Mindray DPM 1 Manuel de l'opérateur Page 1

Naviguer en ligne ou télécharger Manuel de l'opérateur pour Mesure, test et contrôle Mindray DPM 1. Mindray DPM 1 Operator`s manual Manuel d'utilisatio

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 274
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs

Résumé du contenu

Page 1 - Operator’s Manual

DPM 5 Patient Monitor Operator’s Manual

Page 2

2 4 System Menu ... 4-1 4.1 Overvie

Page 3 - Revision History

9-6 9.4 NIBP Recall Select NIBP RECALL >> in SYSTEM MENU. The following window appears. Figure 9-3 NIBP Recall The NIBP RECALL window shows

Page 4

9-7 9.5 Alarm Event Recall Select ALARM RECALL >> in SYSTEM MENU. The following menu appears. Figure 9-4 Alarm Recall Condition Selections

Page 5 - Exemptions

9-8 5. Parameter values at the time of the alarm event. 6. Two waveforms at the time of the alarm event. You can set the waveform length by select

Page 6 - Return Policy

9-9 9.6 Non-Volatile Data Storage To avoid losing patient’s data when the monitor is powered off intentionally or accidentally, this monitor can re

Page 7 - Contact Information

9-10 FOR YOUR NOTES

Page 8 - Preface

10-1 10 ECG/RESP Monitoring 10.1 Overview Two algorithm packages are optional for ECG monitoring: 1. Mindray DS algorithm package This package

Page 9 - Contents

10-2 10.2.2 Electrode Placement WARNING z Use only the specified ECG cable for monitoring. z When applying electrodes or connecting cables, make

Page 10

10-3 3-Leadwire Electrode Placement Following is the configuration per European standard when using three leadwires:  R (right arm) electrode: ne

Page 11

10-4 5-Leadwire Electrode Placement Following is the configuration per American standard when using five leadwires: Figure 10-2 5-Leadwire Electr

Page 12

10-5 When attaching the chest electrode to the back of a patient, place it at one the following sites:  V7: On the 5th intercostal space at the

Page 13

3 6.2.1 Visual Alarms ... 6-2 6.2.2 Audible alarms

Page 14

10-6 10-Leadwire Electrode Placement Figure 10-4 10-Leadwire Electrode Placement 12-lead ECG monitoring uses 10 electrodes which are placed on the

Page 15 - 1 Safety

10-7 WARNING z When using Electrosurgery equipment, patient leads should be placed in a position that is equal distance from the Electrosurgery el

Page 16 - 1.1.2 Warnings

10-8 10.3 ECG Display 10.3.1 ECG Waveform In the standard screen, one or two ECG waveform(s) is (are) displayed at the top of the display when LEAD

Page 17 - 1.1.3 Cautions

10-9 WARNING z Only in the DIAGNOSTIC mode will the monitor provide non-processed real signal waveforms. In the MONITOR or SURGERY mode, the ECG w

Page 18 - 1.1.4 Notes

10-10 10.4 ECG Setup Menu To display the ECG SETUP menu, use the knob to select the ECG label in the parameter windows. The options available in th

Page 19 - 1.2 Equipment Symbols

10-11 The options of ECG SETUP menu are listed below. However, your algorithm package may not include some of these options.  HR ALM Heart rate

Page 20

10-12 1. ECG: HR is always calculated from ECG. 2. SpO2: If the ECG signal is seriously interfered, you can select SPO2, which means PR will be der

Page 21 - 2 The Basics

10-13 Other Setup Select OTHER SETUP >> in the ECG SETUP menu. The following menu appears. The options available in the menu vary with each a

Page 22 - 2.1.4 Functions

10-14 When ECG is in NORMAL DISPLAY mode, if the SMART LEAD OFF is configured ON and there is a LEAD OFF in the HR-derived channel, the system will

Page 23 - 2.2 External Appearance

10-15 NOTE z When monitoring a patient with a pacemaker, PACE must be turned ON. Otherwise, the system will count the pacemaker pulse as QRS compl

Page 24 - 2.2.2 Left Side Panel

4 10.4 ECG Setup Menu ... 10-10 10.5 ST Analysis

Page 25 - 2.2.3 Right Side Panel

10-16 NOTE z If the filter method of the ECG waveform is set to a non-diagnostic mode, only the NOTCH option ON is active and the monitor filters

Page 26 - 2.2.4 Rear Panel

10-17  ECG CAL Select this option to begin calibrating the ECG. To stop calibration, select this option again, or change the ECG lead selection

Page 27 - 2.3 Control Panel

10-18 10.5 ST Analysis 10.5.1 Overview  The function of ST analysis is optional.  In the factory configuration, ST analysis is disabled.  Whe

Page 28 - 2.4 Display

10-19  ST ANAL ST analysis ON: Enables the ST analysis; OFF: Disables the ST analysis. NOTE z ST analysis can be carried out only when the filt

Page 29

10-20 10.5.3 Adjusting ST Measurement Point The ST measurement for each beat complex is the ventical difference between two measurement points. The

Page 30 - 2.5 Batteries

10-21 Mortara algorithm package To adjust the points: 1. Select DEF POINT >> in the ST ANALYSIS menu to open the following window. Three ver

Page 31 - 2.5.1 Battery Guidelines

10-22 10.6 Arrhythmia Analysis 10.6.1 Overview The monitor uses the user-selected primary or secondary ECG lead for arrhythmia analysis. In clinica

Page 32 - 2.5.2 Battery Maintenance

10-23 10.6.2 Arrhythimia Options Mortara algorithm package When Mortara algorithm package is configured, your monitor has the following extended ar

Page 33 - 2.6 Wireless Network Card

10-24 10.6.3 Arrhythmia Analysis Menu Select ARR ANALYSIS >> in ECG SETUP menu. The following menu appears. The menu items vary with the algo

Page 34

10-25 10.6.4 Arrhythimia Relearn During ECG monitoring, you may need to start an arrhythmia relearn when a dramatic change in the patient’s ECG pat

Page 35 - 3 Installation

5 12.4 NIBP Setup Menu ... 12-5 12.4.1 Calibrati

Page 36 - 3.1 Unpacking and Checking

10-26 Figure 10-17 Arrhythmia Alarm Setup (Mortara algorithm package) In the menu, the ALM field indicates the alarm on/off status, REC indicates t

Page 37 - 3.4 Installation Method

10-27 10.6.6 Arr Threshold Select ARR THRESHOLD >> in the ECG SETUP menu and the following menu appears. You can change the threshold setting

Page 38 - 3.4.3 Installing the Battery

10-28 10.6.7 Arrhythmia Recall Select ARR RECALL >> in the ARR ANALYSIS menu. The following menu appears. You can review any stored arrhythmi

Page 39 - 3.4.6 Auxiliary Output Port

10-29 Figure 10-20 Arrhythmia waveform Review You can perform the following operations:  UP-DOWN This option allows you to page up and down to r

Page 40 - 3.6 Powering off the Monitor

10-30 10.7 12-Lead 10.7.1 12-Lead Monitoring Display You can enter the 12-lead monitoring display if your monitor is configured with the 12-lead mo

Page 41 - 4 System Menu

10-31 In the 12-Lead Monitoring Display:  Labels except the ECG-related ones (ECG, ST, GAIN, RHYTHM) are all disabled.  The REC button on the c

Page 42 - 4.2 Patient Setup

10-32  ST ALM ST segment alarm ON: If the measured ST numerics exceed the alarm limit, the monitor gives alarm indications and saves the alarm;

Page 43 - 4.2.1 Admit Patient

10-33 10.8 RESP Monitoring 10.8.1 Overview Respiration is detected by measuring thoracic impedance. The monitor measures the change of the impedane

Page 44 - Setting Patient Information

10-34 10.8.2 Electrode Placement Since the same electrodes are used for ECG and respiration monitoring, the electrode placement is very important.

Page 45 - 4.2.5 Discharge Patient

10-35 10.8.3 Respiration Setup Selecting the RESP label on the screen opens the following menu. Figure 10-25 RESP Setup Menu In this menu, you can

Page 46 - 4.3 Configuration Setup

6 17 Maintenance ... 17-1 17.1 Inspect

Page 47 - 4.4 System Setup

10-36 For different patient types, the upper/lower limits of the respiration rate alarm may vary in the following range. Patient type Upper Limit

Page 48 - 4.4.2 Alarm Setup

10-37 10.9 Maintenance and Cleaning WARNING z Before cleaning the ECG cable, be sure to disconnect the monitor from the ECG cable, or shut down th

Page 49

10-38 FOR YOUR NOTES

Page 50 - 4.4.3 Time Setup

11-1 11 SpO2 Monitoring 11.1 Overview The monitor measures the patients’ SpO2 (oxygen saturation) and displays: 1. Pulse rate (PR) value in the E

Page 51 - 4.4.4 Recorder Setup

11-2 PITCH TONE The PITCH TONE function refers to the monitor’s capability to vary the pitch of the heart rate tone or pulse rate tone with the cha

Page 52 - 4.4.5 Data Output

11-3 11.2.2 Precautions WARNING z The SpO2 value might be overestimated in the presence of Hb-CO, Met-Hb or dye dilution chemicals. z Check if th

Page 53 - 4.4.6 Analog Output

11-4 NOTE z Place the SpO2 sensor cable at the backside of the patient hand. Make sure the fingernail is just opposite to the light emitted from th

Page 54 - 4.4.7 Module Setup

11-5 Figure 11-5 Neonate Sensor Placement (3) NOTE z If the sensor cannot be positioned accurately to the part to be measured, it may result in i

Page 55 - 4.4.8 Trace Setup

11-6 11.2.4 Measurement Limitations If the accuracy of any measurement does not seem reasonable, first check the patient’s vital signs by an altern

Page 56 - 4.4.9 Mark Event

11-7 11.2.5 SpO2 Setup Menu Selecting the SPO2 label in the parameter window opens the following menu. Figure 11-6 SpO2 Setup Menu You can perform

Page 57 - 4.5 Selection Setup

1-1 1 Safety 1.1 Safety Information The safety statements presented in this chapter refer to the basic safety information that the operator of the

Page 58 - 4.6 Monitor Version

11-8 SpO2 and PR alarm limits: Parameter Upper Limit Lower Limit Step SpO2 (lower limit+1) ~100 0~(upper limit-1) 1 PR (lower limit+2) ~254 0~

Page 59 - 4.7 Maintenance

11-9  CONFIG >> You can select this option to access the SPO2 CONFIG menu, in which you may select FACTORY CONFIG or USER CONFIG. After fin

Page 60

11-10  Automatic Calibration Because light absorption by hemoglobin is wavelength dependent and because the mean wavelength of LEDs varies, a mon

Page 61

11-11 11.3.2 Precautions WARNING z Pulse oximeter can overestimate the SpO2 value in the presence of Hb-CO, Met-Hb or dye dilution chemicals. z E

Page 62 - 4.7.1 Alarm Setup

11-12 11.3.4 Measurement Limitations If the accuracy of any measurement does not seem reasonable, first check the patient’s vital signs by an alter

Page 63 - 4.7.2 IP Address Setup

11-13 11.3.5 SpO2 Setup Menu Selecting the SPO2 label in the parameter window opens the following menu. Figure 11-8 SpO2 Setup Menu You can per

Page 64

11-14 SpO2 and PR alarm limits: Parameter Upper Limit Lower Limit Step SpO2 (lower limit+1) ~100 0~(upper limit-1) 1 PR (lower limit+2) ~254 0

Page 65 - 4.7.4 Nurse Call Setup

11-15 11.3.6 Accessories If your monitor is equipped with a Nellcor SpO2 module, use only Nellcor oximetry sensors for SpO2 measurements. Other sen

Page 67 - 4.7.8 Monitor Status

12-1 12 NIBP Monitoring 12.1 Overview The Non-invasive Blood Pressure (NIBP) module measures blood pressure using the oscillometric method. This m

Page 68 - 4.8 DEMO Function

1-2 1.1.1 Dangers There are no dangers that refer to the product in general. Specific “Danger” statements may be given in the respective sections o

Page 69 - 5 Face Selection

12-2 If a set of measured results appears in grey, it indicates this measurement is performed at least 1 hour ago. 12.2 Monitoring Procedure WARN

Page 70 - 5.2 Trend Screen

12-3 5. Make sure that the cuff edge falls within the range of the <-> mark. If it does not, use a larger or smaller cuff that will fit bette

Page 71 - 5.3 OxyCRG Screen

12-4 3. To start a continuous NIBP measurement Selecting CONTINUAL in NIBP SETUP menu starts a continuous NIBP measurement. The monitor continue

Page 72 - 5.4 Viewbed Screen

12-5 12.4 NIBP Setup Menu Selecting the NIBP label in the parameter area opens the following menu. Figure 12-2 NIBP Setup Menu You can perform th

Page 73 - Automatic Selection

12-6  SYS ALM HI Determines the upper limit of the systolic pressure.  SYS ALM LO Determines the lower limit of the systolic pressure.  MEA

Page 74 - 5.5 Large Font Screen

12-7 NS NM ND PR DATE TIME 108 84 70 68 2006-6-12 16:06:00 108 80 70 67 2006-6-12 16:11:00 108 80 70 72 2006-6-12 16:16:00  UNIT Opt

Page 75 - 5.6 Standby Mode

12-8 Figure 12-3 NIBP Calibration NOTE z The calibration of the NIBP measurement should be performed every two years or performed according to t

Page 76

12-9 5. If no information appears on the bottom of the NIBP parameter area, it indicates the airway is in good condition and an air leak does not e

Page 77 - 6 Alarms

12-10 Figure 12-5 Replacing the Rubber Bladder To replace the rubber bladder in the cuff: 1. Place the bladder on top of the cuff so the rubber

Page 78 - 6.2 Alarm Modes

13-1 13 TEMP Monitoring 13.1 Overview The monitor is able to use two temperature probes simultaneously, to obtain two temperature values and the d

Page 79 - 6.2.3 Alarm Messages

1-3 1.1.3 Cautions CAUTION z To ensure patient safety, use only parts and accessories specified in this manual. z Remove the battery from the pat

Page 80 - 6.3 Alarm Statuses

13-2 13.2 Measurement Procedure To measure the temperature of a patient, 1. If a disposable temperature probe is used, plug the temperature probe

Page 81 - 6.3.4 Alarm Sound Off

13-3 13.3 TEMP Setup Menu Selecting the TEMP label in the parameter window opens the following menu. Figure 13-2 TEMP Setup Menu You can perform

Page 82 - 6.4 Latching Alarms

13-4 between channel 1 and channel 2.  TEMP UNIT Options: ℃ and ℉  CONFIG >> You can select this option to access the TEMP CONFIG me

Page 83 - 6.5 Clearing Alarms

14-5 14 IBP Monitoring 14.1 Overview The monitor provides two channels to measure the invasive blood pressure (IBP, including diastolic, systolic

Page 84 - 6.7 When an Alarm Occurs

14-6 14.2 Precautions WARNING z Use only the IBP transducer specified in this operation manual. Disposable IBP transducers or domes should not be

Page 85 - 7 Freezing Waveforms

14-7 14.3 Monitoring Procedure 1. Plug the pressure cable into the IBP connector on the monitor and power on the monitor. 2. Prepare the pressure

Page 86 - 7.3 FROZEN Menu

14-8 14.4 IBP Menu Selecting the IBP (1,2) label in the parameter window opens the following menu. Figure 14-4 IBP (1,2) Select Menu 14.4.1 IB

Page 87 - 7.4 Waveform Recall

14-9 You can perform the following settings in this menu.  ALM IBP alarm on/off status ON: When a IBP alarm occurs, the monitor gives alarm

Page 88

14-10 Figure 14-6 IBP Alarm Setup Menu IBP alarm limits Pressure Label Upper Limit (mmHg) Lower Limit (mmHg) Step (mmHg) ART (lower limit+2) ~30

Page 89 - 8 Recording

14-11 1. HI: Determines the high scale. The reference scale changes when the high scale is changed. 2. LO: Determines the low scale. The referenc

Page 90 - OxyCRG recording

1-4 1.1.4 Notes NOTE z Keep this manual close to the patient monitor so that it can be obtained conveniently when necessary. z This patient monit

Page 91 - 8.3 Recorder Operations

14-12 14.4.2 IBP Pressure Zero Menu You can select IBP PRESURE ZERO in IBP (1,2) SELECT menu to open the following menu. Figure 14-9 IBP Pressure

Page 92 - Alarm event recording

14-13 NOTE z To ensure correct measurement, zero the transducer before it is used to zero the monitor. z Position the transducer at the same leve

Page 93

14-14 14.5 Maintenance and Cleaning WARNING z Before cleaning the transducer, make sure the transducer is disconnected from the monitor, or the m

Page 94 - Removing the Paper Jam

14-15 Sterilization  Chemical Solution Sterilization After finishing the cleaning, select an effective sterilant for chemical solution sterilizat

Page 95 - 9 Recall

14-16 FORYOURNOTES

Page 96 - 9.2 Trend Graph Recall

15-1 15 CO2 Monitoring 15.1 Overview The monitor is able to measure the CO2 pressure of the patient airway, and displays the CO2 waveform in the w

Page 97

15-2 15.2 Mindray DS CO2 Module NOTE z This section is only applicable to a monitor equipped with a Mindray DS CO2 module. 15.2.1 Principles o

Page 98 - 9.3 Trend Table Recall

15-3 15.2.2 Preparations for CO2 Measurement 1. Plug the water trap into its receptacle before the measurement. 2. Open the CO2 SETUP menu and se

Page 99

15-4 The calculation formulas in the above two compensations are as follows: ATPD: PCO2 (mmHg)= CO2 (vol%)× Pamp/100 BTPS: PCO2 (mmHg)= CO2 (vol%)×

Page 100 - 9.4 NIBP Recall

15-5 You can perform the following settings in this menu.  ALM Alarm on/off status ON: When a CO2 alarm occurs, the monitor gives alarm indicatio

Page 101 - 9.5 Alarm Event Recall

1-5 1.2 Equipment Symbols NOTE z Some symbols may not appear on your equipment. Attention: Consult accompanying documents (this manual). Powe

Page 102

15-6 Alarm limits: Parameter Upper Limit Lower Limit Step CO2 (lower limit+2) ~100 0~(upper limit-2) 1 INS 0~100 / 1 AwRR (lower limit+2) ~150

Page 103

15-7 Other Setup Selecting OTHER SETUP option opens the following menu. Figure 15-4 CO2 Other Setup Menu You can perform the following settings i

Page 104 - FOR YOUR NOTES

15-8 When BTPS is turned ON, the system performs BTPS compensation automatically; when BTPS is turned OFF, the BTPS compensation is disabled. When m

Page 105 - 10 ECG/RESP Monitoring

15-9 15.2.4 CO2 User Maintain Menu Select CO2 USER MAINTAIN >> in USER MAINTAIN menu. The following menu appears. Figure 15-5 CO2 User Main

Page 106 - WARNING

15-10  CONFIRM CAL This option allows you to confirm the calibration of the CO2 module. Calibration Procedure 1. Verify the module enters th

Page 107

15-11 15.3 Oridion CO2 Module NOTE z This section is applicable to a monitor equipped with the Oridion CO2 module only. z When the sample flow i

Page 108

15-12 15.3.2 Preparations for CO2 Measurement 1. Plug the sampling line into its receptacle before the measurement. 2. Open the CO2 SETUP menu

Page 109

15-13 15.3.3 CO2 Setup Menu Selecting the CO2 label in the parameter window opens the following menu. Figure 15-8 CO2 Setup Menu You can perform t

Page 110

15-14  AWRR ALM LO Lower alarm limit of AwRR  APNEA ALM Determines the apnea alarm delay. If the apnea of the patient exceeds the preset apnea

Page 111

15-15 You can perform the following settings in this menu.  MAX HOLD If Max Hold is configured to 10 sec or 20 sec, the etCO2 numeric shows the

Page 113 - 10.3.2 ECG Parameters

1-6 Gas outlet ESD sensitivity Manufacture date Serial number European community representative This mark means that this device is fully in c

Page 114 - 10.4 ECG Setup Menu

15-16 15.3.4 CO2 User Maintain Menu Select CO2 USER MAINTAIN >> in USER MAINTAIN menu. The following menu appears. Figure 15-10 CO2 User Mai

Page 115

15-17 Calibration Procedure 1. Verify the module enters the Full Accuracy Mode. 2. Connect a gas bottle with the CO2 sensor connector on the moni

Page 116

15-18 15.3.6 Oridion Information This trademark is registered in Israel, Japan, Germany and US. Oridion Patents This device and the CO2 sampli

Page 117 - Other Setup

16-1 16 Anesthesia Gas Monitoring 16.1 Overview The anesthesia gas (AG) monitoring can be used for measuring the anesthesia gas and respiration ga

Page 118

16-2 Parameters:  CO2: Carbon dioxide  N2O: Nitrous oxide (laughing gas)  O2: Oxygen  AwRR: Airway respiration rate (respiration per m

Page 119

16-3 16.2 Measurement Principles and Procedure Measurement principle of AGs The AG concentration is measured based on the rationale that the AGs h

Page 120

16-4 WARNING z Ensure the compactness of the connection when performing gas measurements. Any leakage in the system will lead to incorrect reading

Page 121

16-5 16.3 MAC Minimum alveolar concentration (MAC) is the minimum concentration of the agent in the alveoli. It is a basic index to indicate the de

Page 122 - 10.5 ST Analysis

16-6 16.4 AG Setup Menu Select the GAS label in the parameter window. The AG SETUP menu appears. Figure 16-4 AG SETUP menu In this menu, you can s

Page 123

16-7  O2 COMPEN The options include: OFF, 30%, 40%, 50%, 60%, 70%, 80%, 90%, and 100%. When the O2 concentration exceeds 60% and it is not being

Page 124

2-1 2 The Basics 2.1 Monitor Description This monitor integrates the functions of parameter measurement, waveform monitoring, freezing, and record

Page 125 - Mortara algorithm package

16-8  EtCO2 ALM LO Set the lower limit of EtCO2 which triggers the alarm  FiCO2 ALM HI Set the upper limit of FiCO2 which triggers the alarm 

Page 126 - 10.6 Arrhythmia Analysis

16-9 Adjust Wave Amplitude Menu Select ADJUST WAVE AMP>> in AG SETUP menu. The following menu appears. Figure 16-7 ADJUST WAVE AMP menu In t

Page 127 - 10.6.2 Arrhythimia Options

16-10 16.5 Calibrating AG Calibrate the AG module every year or when the measured value has a great deviation. Tools required:  Gas bottle, with

Page 128

16-11 8. If the calibration is finished successfully, the message CALIBRATION COMPLETED! is displayed. If the calibration failed, the message AG CA

Page 130

17-1 17 Maintenance WARNING z Failure on the part of the responsible hospital or institution employing the use of the monitoring equipment to imp

Page 131 - 10.6.6 Arr Threshold

17-2 17.2 Cleaning WARNING z Be sure to shut down the system and disconnect all power cords from the outlet before cleaning the equipment. Your e

Page 132 - 10.6.7 Arrhythmia Recall

17-3 17.3 Disinfection and Sterilization Sterilization or disinfection may cause damage to the equipment. We recommend the sterilization and disinf

Page 133

17-4 FOR YOUR NOTES

Page 134 - 10.7 12-Lead

18-1 18 Accessories WARNING z Use accessories specified in this chapter. Using other accessories may cause damage to the patient monitor or not m

Page 135

2-2 2.1.4 Functions This monitor is capable of monitoring the following parameters.  ECG  Heart rate (HR)  ECG waveform(s)  ST segment anal

Page 136

18-2 18.1 ECG Accessories ECG Electrodes Model Quantity Patient Category Part No. 210 10 pieces Adult 0010-10-123042258-3 3 pieces Neonate 900E-

Page 137 - 10.8 RESP Monitoring

18-3 18.2 SpO2 Accessories Extension Cable Module type Part No. Mindray DS SpO2 Module 0010-20-42594 SpO2 Sensors The SpO2 sensor material that

Page 138 - 10.8.2 Electrode Placement

18-4 18.3 NIBP Accessories Tubing Type Patient Category Part No. Adult, pediatric 509B-30-06259 Reusable Neonate 509B-30-06260 Reusable Cuff Mo

Page 139 - 10.8.3 Respiration Setup

18-5 18.4 TEMP Accessories Extension Cable Type Model Temp probe Part No. Reusable MR420 MR411, MR412 0011-30-90444 Temp Probes Type Model

Page 140

18-6 Manufacturer Accessory Smith Medical (Medex) MX961Z14 trunk cable MX960 transducer MX9605A monitoring kit MX960E6441 transducer seat MX261

Page 141

18-7 18.6 CO2 Accessories 18.6.1 Mindray DS CO2 Accessories Material Patient Category Remark Part No. DRYLINE Watertrap Adult, pediatric 9200-

Page 142

18-8 18.8 Others Model Part No. M05-010001-06 (black) Lithium battery 0146-00-0099 (gray) 023-000034-00 CF storage card 0000-10-10798 Data ou

Page 143 - Monitoring

A-1 A Product Specifications A.1 Safety Classifications Type of protection against electric shock Class I with internal electric power supply. Whe

Page 144 - Module

A-2 A.3 Power Source Specifications AC Power Supply Specifications Input voltage 100 to 240 V~ Current 1.4A to 0.6A Frequency 50/60 Hz Fuse T 3

Page 145 - 11.2.3 Monitoring Procedure

A-3 A.4 Hardware Specifications Physical Size 318 × 270 × 137mm (width×height×depth) Weight Different due to different configurations Standard con

Page 146 - Neonate Sensor Placement

2-3 2.2 External Appearance 2.2.1 Front Panel Figure 2-1 Front Panel This monitor is designed to comply with the requirements of relative interna

Page 147

A-4 A.5 Data Storage Trend data Long trend: 96 hours, resolution 1min, 5 min or 10 min. Short trend: 1 hour, resolution 1 s or 5 s. Alarm events 70

Page 148

A-5 Conducting resistance < 1Ω Isolation voltage > 1500 VAC Signal type Normally open or normally closed, selectable Defibrillator synchron

Page 149 - Setup Menu

A-6 Input signal range ±8mV (peak-to-peak value) Accuracy of input signal reproduction Methods A and D were used to establish overall system error

Page 150

A-7 Response time of tachycardia alarm When tested in accordance with ANSI/AAMI EC13 Section 4.1.2.1 g, the response time is as follows: Figure 4ah

Page 151 - 11.3 Nellcor SpO

A-8 A.8.2 Mortara Software Package Lead naming style AHA, EURO Lead fault The lead resistance is no greater than 51 kΩ and it is in parallel with

Page 152

A-9 The monitor complies with the test method in EC13 Section 5.2.9.14. HR Measurement range Neonate: Pediatric: Adult: 15 to 350 bpm 15 to 350 bpm

Page 153 - 11.3.3 Monitoring Procedure

A-10 Amplitude: Width: Rise time: ±4 to ±700 mV (3/5-lead) ±2 to ±700 mV (12-lead) 0.1 to 2 ms 10 to 100 µs When tested in accordance with the ANSI

Page 154

A-11 A.10 SpO2 Specifications A.10.1 Mindray DS SpO2 Module All SpO2 sensors specified in the section Mindray DS SpO2 Accessories meets the followi

Page 155 - 11.3.5 SpO

A-12 A.10.2 Nellcor SpO2 Module All SpO2 sensors specified in the section Nellcor SpO2 Accessories meets the following specifications when used wit

Page 156

A-13 A.11 NIBP Specifications Measurement technique Auto oscillation Displayed parameters Systolic pressure, diastolic pressure, mean pressure an

Page 157 - 11.3.7 Nellcor Information

2-4 2.2.2 Left Side Panel Your monitor may not have all connectors as shown below. Figure 2-2 Left Side Panel 1. T1: Temperature probe connec

Page 158 - FORYOURNOTES

A-14 A.12 TEMP Specifications Number of channels 2 Displayed parameters T1, T2 and TD Measurement range 0 to 50°C (32 to 122°F) Resolution 0

Page 159 - 12 NIBP Monitoring

A-15 Resolution 1 bpm Pressure transducer Excitement voltage 5 VDC, ±2% Sensitivity 5 uV/V/mmHg Impedance range 300 to 3000Ω Volume displacement

Page 160 - 12.2 Monitoring Procedure

A-16 <3 s @ 100 ml/min <3.5 s @ 70 ml/min When measured with an adult watertrap and a 2.5m-long adult sampling line: <5 s @ 100 ml/min <

Page 161 - 12.2.2 Operation Guides

A-17 A.15 AG Specifications Measurement technique Infrared absorption Measurement mode Sidestream AG functions Meets requirements of ISO9918, ISO

Page 162 - 12.3 Measurement Limitations

A-18 > 8 Not specified 0 to 1 ±0.15 1 to 5 ±0.2 Enf, Iso, Hal > 5 Not specified 0 to 25 ±1 25 to 80 ±2 O2 (Optional) 80 to 100 ±3 2 to 60 Br

Page 163 - 12.4 NIBP Setup Menu

A-19 Effect of interference gases on AG measurements Quantitive effect(%ABS)3) Gas Concentration(%)CO2 N2O Agent 1) O2 CO2 / / 0.1 0 0.2 N2O /

Page 164

A-20 FOR YOUR NOTES

Page 165 - 12.4.1 Calibration

B-1 B EMC The equipment meets the requirements of EN 60601-1-2. NOTE z Use of accessories, transducers, and cables other than those specified ma

Page 166

B-2 TABLE 2 Guidance and declaration — electromagnetic immunity The equipment is intended for use in the electromagnetic environment specified bel

Page 167

B-3 TABLE 3 Guidance and declaration — electromagnetic immunity The equipment is intended for use in the electromagnetic environment specified bel

Page 168 - CAUTION

2-5 2.2.3 Right Side Panel The right side panel differs with different configurations as shown below. Figure 2-3 Right Side Pane

Page 169 - 13 TEMP Monitoring

B-4 TABLE 4 Recommended separation distances between portable and mobile RF communication and the equipment The equipment is intended for use in an

Page 170 - 13.2 Measurement Procedure

C-1 C Alarm Messages and Prompt Information NOTE z Your patient monitor may not include all the information listed in this section. C.1 Physi

Page 171 - 13.3 TEMP Setup Menu

C-2 PVC 2* The PVC arrhythmia event occurs to the patient. BIGEMINY 2* The bigeminy arrhythmia event occurs to the patient. TRIGEMINY 2* The t

Page 172

C-3 C.2.2 ECG Module Alarm Messages Alarm message A B L Cause Measure ECG LEAD OFF No Yes 3 ECG X LEAD OFF No Yes 3 The ECG lead is not con

Page 173 - 14 IBP Monitoring

C-4 C.2.4 TEMP Module Alarm Messages Alarm message A B L Cause Measure TEMP1 SENSOR OFF No Yes 3 The T1 sensor is not connected correctly t

Page 174 - 14.2 Precautions

C-5 Alarm message A B L Cause Measure NIBP SYSTEM FAILURE No Yes 2 NIBP TIME OUT No Yes 2 MEASURE FAIL No Yes 2 Failure occurs in the pulse mea

Page 175 - 14.3 Monitoring Procedure

C-6 C.2.7 Nellcor SpO2 Module Alarm Messages Alarm message A B L Cause Measure SPO2 SENSOR OFF No Yes 3 The sensor is disconnected from the pa

Page 176 - 14.4 IBP Menu

C-7 C.2.8 IBP Module Alarm Messages Alarm message A B L Cause Measure IBP1 SENSOR OFF No Yes 3 The invasive blood pressure cable of channel

Page 177

C-8 Alarm message A B L Cause Measure CO2 BAROMETRIC TOO HIGH No No 2 The barometric pressure is too high. CO2 BAROMETRIC TOO LOW No No

Page 178

C-9 Alarm message A B L Cause Measure CO2 INIT ERR Yes No 1 The CO2 module is not properly installed, or malfunctions. CO2 COMM STOP No N

Page 179

2-6 2.2.4 Rear Panel Figure 2-4 Rear Panel 1. Fan Vent 2. Speaker holes 3. Mounting holes for support bracket. 4. Network Connector: Standard R

Page 180

C-10 Alarm message A B L Cause Measure CO2 CALIB ERROR No No 2 Including the error messages displayed at the lower part of the menu (like gas e

Page 181

C-11 exceeds 1 second. AG SELF-TEST ERR Yes No 1 AG module fails or fails to communicate with the host. Unplug and re-plug the module and th

Page 182

C-12 C.2.13 Recorder Module Alarm Messages Alarm message A B L Cause Measure RECORDER INIT ERR N Yes No 2 An error occurs during the record

Page 183 - Sterilization

C-13 C.2.14 System Alarm Messages Alarm message A B L Cause Measure REAL CLOCK NEED SET No No 1 The system time is incorrect. Reset the sy

Page 184

C-14 C.3 Prompt Messages Prompt messages Cause Measure ECG1 SIGNAL SATURATION ECG2 SIGNAL SATURATION Signals of abrupt change interfere with the

Page 185

C-15 Prompt messages Cause Measure AG IS STARTING The AG module is starting. Wait for the AG module to finish the startup. AG WARM UP The A

Page 187 - Measurement

D-1 D Symbols and Abbreviations Symbols and abbreviations that you may encounter while reading this manual or using the monitor are listed below w

Page 188 - 15.2.3 CO

D-2 mmHg millimeters of mercury ms millisecond mV millivolt mW milliwatt nm nanometer ppm part per million s second V volt VA volt amper

Page 189

D-3 D.2 Abbreviations AAMI Association for Advancement of Medical Instrumentation AC alternating current ADT adult AG anaesthesia gas AHA Amer

Page 190

2-7 WARNING z Accessory equipments connected to this patient monitor must be certified according to the respective IEC standards (e.g. IEC 60950 f

Page 191

D-4 ERR error ES electrosuigical ESU electrosuigical unit Et end-tidal EtCO2 end-tidal carbon dioxide EtN2O end-tidal nitrous oxide EtO Et

Page 192

D-5 MAC minimal alveolar concentration MAP mean arterial pressure MDD Medical Device Directive MEAN mean pressure MetHb methemoglobin Mii init

Page 193 - User Maintain Menu

D-6 SYNC synchronization SYS systolic T1 temperature of channel 1 T2 temperature of channel 2 TB temperature of blood TD temperature differenc

Page 195 - 15.3 Oridion CO

Mindray DS USA, Inc.800 MacArthur Blvd.Mahwah, New Jersey 07430USATel:1.800.288.2121Tel:1.201.995.8000www.mindray.comP/

Page 196 - 15.3.2 Preparations for CO

2-8 10. Control Knob The main operator control is the control knob. The control knob rotates in either direction to highlight parameter labels an

Page 197 - 15.3.3 CO

2-9 2. System time The system time of the monitor is displayed in two lines. The time format can be set in the TIME SETUP menu. 3. Technical ala

Page 198

I Revision History This manual has a revision number. This revision number changes whenever the manual is updated due to software or technical spec

Page 199

2-10 2.5 Batteries This monitor is designed to operate on battery (sealed lead-acid or Lithium Ion battery) power during intra-hospital patient tra

Page 200 - 15.3.4 CO

2-11 2.5.1 Battery Guidelines Life expectancy of a battery depends on how frequent and how long it is used. For a properly maintained and stored le

Page 201

2-12 2.5.2 Battery Maintenance Conditioning a Battery A battery should be conditioned before it is used for the first time. A battery conditioning

Page 202 - 15.3.6 Oridion Information

2-13 NOTE z The battery might be damaged or malfunctioned if its operating time is too short after being fully charged. The operating time depends

Page 203 - 16 Anesthesia Gas Monitoring

2-14 In front of the wireless network card are three indicators that indicating its working status:  POWER indicator  OFF: The wireless network

Page 204

3-1 3 Installation WARNING z The installation of the monitor must be carried out by personnel authorized by us. The software copyright of the mon

Page 205 - Measurement principle of O

3-2 3.1 Unpacking and Checking Before unpacking, examine the packing case carefully for signs of damage. If any damage is detected, contact the car

Page 206

3-3 3.3 Power Source Requirements The power applied to the monitor must meet the requirements specified in this manual. WARNING z Make sure that

Page 207 - MAC =+ + =

3-4 3.4.3 Installing the Battery If the monitor is to be powered by the internal battery, install the battery following the steps as below: 1. Ope

Page 208 - 16.4 AG Setup Menu

3-5 3.4.6 Auxiliary Output Port The auxiliary output port can be used to generate analog signals, nurse call signals or defibrillator synchronizati

Page 209 - Alarm Setup Menu

II Intellectual Property Statement Mindray DS USA, Inc. (hereinafter called Mindray DS) owns the intellectual property rights to this product and t

Page 210

3-6 3.4.8 Equipotential Grounding When other equipments are used together with the monitor, a grounding cable should be used to connect the equipot

Page 211 - Configuration Menu

4-1 4 System Menu 4.1 Overview This chapter only gives introduction to the system menu. Other menus will be described in the following chapters.

Page 212 - 16.5 Calibrating AG

4-2 4.2 Patient Setup Select PATIENT SETUP>> in SYSTEM MENU. The following menu appears. Figure 4-2 Patient Setup Menu This menu displays th

Page 213 - 16.6.1 Occlusion handling

4-3 4.2.1 Admit Patient To admit a new patient, please follow this procedure: 1. Select ADMIT PATIENT in PATIENT SETUP menu. 2. Select YES in the

Page 214

4-4 NOTE z If the PAT NO or NAME has not been input, “PATI. INFO. IMCMP” will be displayed in the patient information area. 4. Select OK button,

Page 215 - 17 Maintenance

4-5 4.2.2 Quick Admit Patient 1. Select QUICK ADMIT PATIENT in PATIENT SETUP menu. 2. Select YES in the pop-up CONFIRM TO CLEAR THE DATA menu. 3.

Page 216 - 17.2 Cleaning

4-6 4.3 Configuration Setup Select CONFIG >> in the SYSTEM MENU. The following menu appears. Figure 4-5 Configuration menu If you have chang

Page 217

4-7 4.4 System Setup Select SYSTEM SETUP>> in SYSTEM MENU. The following menu appears. Figure 4-6 System Setup SYSTEM SETUP menu contains t

Page 218

4-8 4.4.1 Face Select Select FACE SELECT>> in SYSTEM SETUP menu. The following menu appears. Figure 4-7 Face Select In the FACE SELECT menu

Page 219 - 18 Accessories

4-9 You can perform the following settings in the menu above:  ALM SEL Alarm selection Options: COMMON ALM SETUP, XX ALM SETUP; (XX refers to phy

Page 220 - 18.1 ECG Accessories

III Manufacturer’s Responsibility All information contained in this manual is believed to be correct. Mindray DS shall not be liable for errors con

Page 221 - Accessories

4-10 4.4.3 Time Setup Select TIME SETUP>> in SYSTEM SETUP menu. The following menu appears. Figure 4-9 Time Setup With the control knob, y

Page 222 - 18.3 NIBP Accessories

4-11 4.4.4 Recorder Setup Select RECORD>> in SYSTEM SETUP menu. The following menu appears. Figure 4-10 Recorder Setup In this menu, you ca

Page 223 - 18.5 IBP Accessories

4-12  RT REC TIME Real-time recording time Options: CONTINUAL and 8s.  TIMING REC TIME Timing recording time The interval between automatic rec

Page 224

4-13 Output Procedure 1. Disconnect all patient cables connected to the monitor. 2. Verify the monitor is connected to the PC and the PC is runni

Page 225 - 18.7 AG Accessories

4-14 NOTE z If DEFIB. SYN or NURSE CALL is selected from the AUX OUTPTU options in the USER MAINTAIN menu, the ANALOG>> option in the SYSTEM

Page 226 - 18.8 Others

4-15 4.4.8 Trace Setup Select TRACE SETUP>> in SYSTEM SETUP menu. The following menu appears. Figure 4-14 Trace Setup This menu allows you t

Page 227 - A Product Specifications

4-16 4.4.9 Mark Event Select MARK EVENT>> in SYSTEM SETUP menu. The following menu appears. Figure 4-16 Mark Event This menu allows you to

Page 228

4-17 4.5 Selection Setup Select SELECTION>> in SYSTEM MENU. The following menu appears. Figure 4-17 Selection Setup You can perform the foll

Page 229 - A.4 Hardware Specifications

4-18 4.6 Monitor Version You can select VERSION>> in the SYSTEM MENU to check the version information as shown below.. However, the VERSION m

Page 230 - A.6 Wireless Network

4-19 You can also see the device version by selecting DEVICE VERSION LIST>> as shown below. However, the DEVICE VERSION LIST menu of you moni

Page 231 - A.8 ECG Specifications

IV Return Policy In the event that it becomes necessary to return a unit to Mindray DS, follow the instructions below. 1. Return authorization. Co

Page 232

4-20 Enter USER KEY, then select CONFIRM button. The following menu appears. Figure 4-22 User Maintain You can perform the following settings:

Page 233

4-21 1. ANALOG OUT (analog output)  If this option is selected, the auxiliary output port will be able to output analog signals, and you can per

Page 234

4-22 NOTE z If you changed the net type and language, the settings take effect only after you restart the monitor. z The setting of the notch fre

Page 235

4-23 NOTE z In the case that alarm volume is set to 0, if you set ALARM SOUND OFF to DISABLE, alarm volume change to the default. WARNING z S

Page 236 - A.9 RESP Specifications

4-24 4.7.3 Self Definition of Color Select COLOR SELF-DEFINE >> in USER MAINTAIN menu. The following menu appears. Figure 4-25 Self-definiti

Page 237 - Specifications

4-25 4.7.4 Nurse Call Setup Select NURSE CALL SETUP >> in USER MAINTAIN menu. The following menu appears. Figure 4-26 Nurse Call Setup You

Page 238 - A.10.2 Nellcor SpO

4-26 Trigger Conditions A nurse call signal will be triggered only if all the following conditions are met: 1. The nurse call function is enable

Page 239 - A.11 NIBP Specifications

4-27 4.7.8 Monitor Status Select STATUS >> in ENTER MAINTAIN PASSWORD menu. The following menu appears. Figure 4-27 Monitor Status  UP-D

Page 240 - A.13 IBP Specifications

4-28 4.8 DEMO Function Select DEMO >> in SYSTEM MENU. The following menu appears. Figure 4-28 Input Demo Key The monitor enters the demonst

Page 241

5-1 5 Face Selection 5.1 Standard Screen You can open the FACE SELECT menu by selecting FACE SELECT >> in SYSTEM SETUP menu. Figure 5-1 Fac

Page 242 - A.14.2 Oridion CO

V Contact Information Manufacturer: Mindray DS USA, Inc. Address: 800 MacArthur Blvd.Mahwah, New Jersey 07430 USA Tel: 1.800.288.2121 1.201.

Page 243 - A.15 AG Specifications

5-2 5.2 Trend Screen To enter the following screen, select TREND SCREEN in FACE SELECT menu and then select EXIT. Figure 5-3 Trend Screen  Trend

Page 244

5-3 5.3 OxyCRG Screen To enter the following screen, select oxyCRG SCREEN in FACE SELECT menu and then select EXIT. Figure 5-4 OxyCRG Screen Oxy

Page 245

5-4 5.4 Viewbed Screen This monitor can view one parameter waveform and measured data from another patient monitor (viewbed monitor) on the same mo

Page 246

5-5 1. Viewbed monitor label The viewbed monitor lable allows you to select the viewbed monitor you want to view. It displays the bed number and

Page 247

5-6 5.5 Large Font Screen To enter the following screen, open FACE SELECT menu, select LARGE FONT SCREEN, and then select EXIT. Figure 5-6 Large F

Page 248

5-7 5.6 Standby Mode During patient transport or temporary departure of a patient, the monitor can be set to STANDBY mode. In this mode, the monito

Page 250 - Pd 3.2=

6-1 6 Alarms 6.1 Overview The monitor gives audible or visual alarms to indicate the medical staff, when a vital sign of the patient appears abnor

Page 251

6-2 6.1.2 Alarm Levels The alarms are divided into three priority levels: high level alarms, medium level alarms and low level alarms. 1. High le

Page 252 - C.2 Technical Alarm Messages

6-3 6.2.2 Audible alarms The monitor uses different alarm tones to indicate different alarm levels.  High level alarm: “DO-DO-DO--DO-DO---

Page 253

VI Preface Manual Purpose This manual contains the instructions necessary to operate the product safely and in accordance with its function and int

Page 254

6-4 6.2.4 Parameter Flashes An alarm is triggered when a patient parameter exceeds the parameter limit. At the same time, the measured parameter va

Page 255 - Module Alarm Messages

6-5 NOTE z When a new parameter module is installed or when a parameter module is turned ON, all the parameter alarms and technical alarms related

Page 256 - C.2.7 Nellcor SpO

6-6 6.3.5 Status Switchover 1. In the Normal status,  Press the SILENCE key for less than 2 seconds to switch to the Alarms Paused status, or 

Page 257 - C.2.10 Mindray DS CO

6-7 6.5 Clearing Alarms Generally the alarm indications of an alarm will automatically be cleared when the alarm condition that triggered the alarm

Page 258

6-8 6.6 Testing Alarms When the monitor starts up, a selftest will be performed automatically. The alarm indicator will flash once in yellow and re

Page 259 - C.2.11 Oridion CO

7-1 7 Freezing Waveforms 7.1 Overview You can freeze the monitored waveforms of a patient as desired and view the waveforms of 40 seconds to gain

Page 260

7-2 7.3 FROZEN Menu The FROZEN menu is displayed at the lower left corner. You can perform the following settings in this menu.  WAVE 1 Waveform

Page 261

7-3 7.4 Waveform Recall 1. In the frozen mode, select the RECALL option in the FROZEN menu, and the option name changes to L-RIGHT. 2. Rotate the

Page 262

7-4 FOR YOUR NOTES

Page 263 - C.2.14 System Alarm Messages

8-1 8 Recording 8.1 Overview A thermal recorder can be equipped with the monitor. The performance of the recorder is described as below.  Record

Page 264 - C.3 Prompt Messages

1 Contents 1 Safety...

Page 265

8-2 Alarm recording Alarm recording is measured parameter recording. When a parameter alarm occurs, the recorder automatically records two waveform

Page 266

8-3 8.3 Recorder Operations Continuous real-time recording 1. Press the RECORD key to start recording. 2. Press the RECORD key again to stop the

Page 267 - D Symbols and Abbreviations

8-4 Trend table recording 1. Select TREND TABLE>> in SYSTEM MENU to open the TREND TABLE window. 2. Select the REC option to start recordin

Page 268

8-5 8.4 Installing Recorder Paper Installing Procedure 1. Press the latch at the upper right of the paper compartment door to releases the door. 2

Page 269 - D.2 Abbreviations

8-6 Removing the Paper Jam If the recorder does not function properly or produces unusual sound, open the recorder door to check for a paper jam. Y

Page 270

9-1 9 Recall 9.1 Overview The monitor is able to store important patient data so that the user can review and record the data as desired.  Tren

Page 271

9-2 9.2 Trend Graph Recall Select TREND GRAPH >> in SYSTEM MENU. The following window appears. 1. Trend cursor 2. Cursor time 3. Trend gra

Page 272

9-3 When “OFF” is selected for “PARA1”, “PARA2” or “PARA3”, its trend graph will disappear, and then the trend graphs of other parameters will in

Page 273

9-4 9.3 Trend Table Recall Select TREND TABLE >> in SYSTEM MENU. The following window appears. Figure 9-2 Trend Table The TIME is displayed

Page 274

9-5  UP-DOWN 1. Rotate the control knob to highlight the UP-DOWN option, and press. 2. If the mark is displayed at the lower right of the TIME

Commentaires sur ces manuels

Pas de commentaire