A-4
A.5 Data Storage
Trend data
Long trend: 96 hours, resolution 1min, 5 min or 10 min.
Short trend: 1 hour, resolution 1 s or 5 s.
Alarm events
70 alarm events and associated waveforms (with user
selectable waveform length 8s, 16 or 32).
NIBP measurements
800 NIBP groups, including systolic pressures, mean pressures,
diastolic pressures and measurement time.
A.6 Wireless Network
Standards IEEE 802.11g, Wi-Fi compatible
Frequenct range 2.412 to 2.462GHz
China America Canada Europe Spain France Japan
1 to 11 10, 11 2
Operating channel
For other country, please refer to your local law.
Safe distance 10 m (a circle centering AP with the diameter of 10 m)
A.7 Signal Output Specifications
Standards
Meets the requirements of EC60601-1 for short-circuit
protection and leakage current
Output impedance 50
ECG analog output
Bandwidth (-3dB; reference
frequency: 10Hz)
Diagnostic mode:
Monitor mode:
Surgery mode:
0.05 to 100 Hz (812A module)
0.05 to 150 Hz (M08A module)
0.5 to 40 Hz
1 to 20 Hz
Maximum propagation delay 25 ms (In DIAGNOSTIC mode, NOTCH is OFF)
Sensitivity 1 V/mV± 5%
PACE rejection/enhancement No pace rejection or enhancement
IBP analog output
Bandwidth 0 to 12.5 Hz (-3 dB, reference frequency: 1 Hz)
Maximum propagation Delay 55 ms (the filter function is disabled)
Sensitivity 1 V/100 mmHg ±5%
Nurse call output
Driver Relay
Electrical specifications 60W, 2A, 36VDC, 25VAC