A-16
<3 s @ 100 ml/min
<3.5 s @ 70 ml/min
When measured with an adult watertrap and a 2.5m-long adult
sampling line:
<5 s @ 100 ml/min
<6.5 s @ 70 ml/min
Apnea alarm delay AwRR: 10 to 40 s
* Conditions for measurements in typical precision:
The measurement is started after the preheating mode of the module;
Ambient pressure: 750 mmHg to 760 mmHg; room temperature: 22℃ to 28℃;
The gas under test is dry, and the balance gas is N2;
The deflation rate is 100 ml/min, the respiration rate is no greater than 50 BrPM, with a
fluctuation less than ±3 BrPM, and the inhale interval/exhale interval is 1:2;
When the working temperature is from 15 to 25 degree, or from 50 to 55 degree, or when the
breath rate is greater than 50Brpm, the measurement precision should meet the requirements of
ISO21647: ±4mmHg (0 to 40mmHg) or ±12% of the reading (41 to 99 mmHg)
A.14.2 Oridion CO
2
Specifications
CO2 measurement range 0 to 99mmHg
Precision*
0 to 38 mmHg:
39 to 99 mmHg:
±2 mmHg
±5% + 0.08% × (reading - 38 mmHg)
Drift meet the requirement of accurancy in 6 hours
Resolution
Waveform:
Value:
0.1 mmHg
1 mmHg
Sample flow rate
7.5
15
50
−
+
ml/min
Initialization time 30 s (typical)
Response time 2.9 s (typical)
Delay time 2.7 s (typical)
AwRR measurement range 0 to 150 BrPM
AwRR measurement
precision
0 to 70 BrPM:
70 to 120 BrPM:
121 to 150 BrPM:
±1 BrPM
±2 BrPM
±3 BrPM
Apnea alarm delay AwRR: 10 to 40 s
* Precision applies for breath rates of up to 80 bpm. For breath rates above 80 bpm, accuracy
complies with EN ISO 21647/ISO 21647/ISO 9918 (4 mmHg or ±12% of reading whichever is
greater) for EtCO2 values exceeding 18 mmHg. To achieve the specified accuracies for breath
rates above 60 breaths/minute, the Microstream® FilterLine H Set for Infant/Neonatal (p/n
006324) must be used.The accuracy specification is maintained to within 4% of the values
indicated in the above table in the presence of interfering gases according to EN ISO
21647/ISO 21647 Section Eleven, Part 101.