Mindray DPM 1 Manuel de l'opérateur Page 189

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 274
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 188
15-5
You can perform the following settings in this menu.
ALM Alarm on/off status
ON: When a CO2 alarm occurs, the monitor gives alarm
indications and stores the alarm;
OFF: When a CO2 alarm occurs, the monitor neither gives
alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the
right of the CO2 label.
ALM LEV Alarm level
Options: HIGH and MED.
ALM REC Alarm recording
ON: When a CO2 alarm occurs, the monitor enables the
recording;
OFF: When a CO2 alarm occurs, the monitor does not
enable the recording.
CO2 ALM HI Upper alarm limit of EtCO2
CO2 ALM LO Lower alarm limit of EtCO2
INS ALM HI Upper alarm limit of InsCO2
AWRR ALM HI Upper alarm limit of AwRR
AWRR ALM LO Lower alarm limit of AwRR
APNEA ALM Determines the apnea alarm delay.
If the apnea of the patient exceeds the preset apnea alarm
delay, the monitor triggers an alarm and gives the " CO2
APNEA " alarm message.
Options: 10S, 15S, 20S, 25S, 30S, 35S and 40S.
SWEEP Waveform speed
Options: 6.25, 12.5 and 25.0.
Unit: mm/s.
UNIT Options: mmHg, kPa and %.
WAVE SCALE Options: HIGH and LOW.
This option allows you to adjust the amplitude of the CO2
waveform.
WORK MODE Options: MEASURE and STANDBY.
To start the CO2 monitoring, select the MEASURE mode.
In the STANDBY mode, the air pump and infrared ray
source of the CO2 module are disabled. Hence, decreases
the power consumption and prolongs the operating life of
infrared ray source and the whole CO2 module.
Vue de la page 188
1 2 ... 184 185 186 187 188 189 190 191 192 193 194 ... 273 274

Commentaires sur ces manuels

Pas de commentaire