10-14
When ECG is in NORMAL DISPLAY mode, if the SMART LEAD OFF is configured ON and
there is a LEAD OFF in the HR-derived channel, the system will automatically switch to another
lead to restore the display of the ECG waveform in the HR-derived channel. When the LEAD
OFF condition is corrected, the leads are automatically switched back. However, if you manually
switch the HR-derived channel, change the ECG display mode, or turn off SMART LEAD OFF
before the LEAD OFF condition is corrected, the leads will not be switched back.
Switching rules for SMART LEAD OFF:
LEAD OFF condition HR-derived channel automatically switches to
RA OFF III
LL OFF I
LA OFF II
V OFF II
RL OFF None
Any two out of LL, RA and LA fall off. None
LL, RA or LA, and V fall off together Same as LL, RA or LA falls off.
NOTE
z The SMART LEAD OFF function is available only when the LEAD TYPE is 5
LEADS or 12 LEADS.
z When the HR CHANNEL is set to AUTO, SMART LEAD OFF does not work as
long as there is ECG waveform displayed in either channel 1 or 2.
z When you perform ECG calibration, SMART LEAD OFF does not work.
BEAT VOL Beat volume
Range: 0-10. 0 indicates disabled and 10 indicates the
maximum volume.
PACE ON: When ON is selected, a detected pacemaker signal is
indicated by a "|" symbol above the ECG waveform.
OFF: When OFF is selected, the pacemaker analysis is
disabled.