Mindray DPM 1 Manuel de l'opérateur Page 213

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 274
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 212
16-11
8. If the calibration is finished successfully, the message CALIBRATION COMPLETED! is
displayed. If the calibration failed, the message AG CAL. FAILED is displayed. Perform
another calibration.
9. Select EXIT to exit the current menu.
16.6 Maintenance and Cleaning
16.6.1 Occlusion handling
If the AG module passage is occluded, the screen will prompt “AG OCCLUSION”. Following
are a few examples of occlusion, which you may remove one by one until this message
disappears.
Entrance occlusion
If the part at the entrance such as the filter, sample line or airway adapter is occluded by the
condensed water, the system will prompt on the screen that the airway is occluded.
The optimal method to remove clogs of this kind is:
Check the filter for clogs. If it is clogged, replace the bacteria filter at the entrance.
Check the sample pipe for clogs or knots. Replace it if necessary.
Check the airway adapter for water. If necessary, drain the water and install the adapter
again.
Internal Occlusion
If the interior of the AG module is contaminated by the condensed water, the system will prompt
on the screen that the airway is occluded.
The optimal method to remove clogs of this kind is:
1. Check, as usual, the entrance or the exit for clogs, and remove them if any.
2. If the occlusion still exists after step 1, check for the existence of interior occlusion. In this
case, contact our Customer Service.
16.6.2 Removing Exhaust Gases from the System
WARNING
z Anesthetics: When using the AG measurement on patients who are receiving or
have recently received anesthetics, connect the outlet to a scavenging system, or
to the anesthesia machine/ventilator, to avoid exposing medical staff to
anesthetics.
To remove the sample gas to a scavenging system, connect an exhaust tube to the gas outlet
connector of the module.
Vue de la page 212
1 2 ... 208 209 210 211 212 213 214 215 216 217 218 ... 273 274

Commentaires sur ces manuels

Pas de commentaire