16-4
WARNING
z Ensure the compactness of the connection when performing gas measurements.
Any leakage in the system will lead to incorrect readings because this leakage
will make the surrounding environmental air mix up with the patient gas.
z The water trap is used for collecting water condensates to protect the module
from the ingress of water; when the collected water in the water trap reaches a
specific volume, it must be removed, thus to avoid clogging in the gas system.
z The water trap has filter materials inside, which protect the module from being
contaminated by bacterium, liquid, or patient secretion. When the water trap
works for a long time, dusts or other foreign bodies may block the filter
materials, or even result in system clogging. In this case, replace the water trap.
The recommended replacement period is 2 months.
z Only gas sampling hoses recommended for the patient monitor are permitted.
Otherwise, the performance and reliability of the AG module may be decreased.
z In case a gas sampling hose is knotted, disuse it. Because the knot may cause
clogging or leakage.
z When an inspiratory anesthetic is used, use an exhaust hose to connect the outlet
of the AG module to the exhaust disposal system of the hospital or to the
anesthetic/respiration machine.
z Use of high frequency electrosurgical equipment may increase the risk of burns.
Do not use antistatic or conductive masks or breathing tubes.