10-16
NOTE
z If the filter method of the ECG waveform is set to a non-diagnostic mode, only the
NOTCH option ON is active and the monitor filters the signals of the power line
frequency.
CASCADE ECG cascade
ON: The monitor displays the waveform of each channel in
two lines.
OFF: The monitor displays the waveform of each channel in
one line.
The CASCADE can be turned ON only if the monitor is set
to NORMAL DISPLAY mode and the SCAN TYPE is set
to REFRESH.
DEFIB SYNC Defibrillator synchronization
ON: The function of defibrillation is enabled;
OFF: The function of defibrillation is disabled.
If the function of defibrillator synchronization is enabled and the ventricular defibrillation is
applied, 100ms/+5V defibrillator synchronization signals will be output through the auxiliary
output connector to the defibrillation equipment.
When the defibrillator synchronization is enabled, the “Defib Sync ON!” message is displayed in
the prompt information area, at the lower left corner of the screen. This message disappears after
10 seconds. Besides, the technical alarms area also shows the “DEFIB SYNC ON” message.
NOTE
z Defibrillator synchronization can be enabled only if the AUX OUTPUT in the
USER MAINTAIN menu is set to DEFIB. SYN. Otherwise, this function is
disabled.
z When the monitor is started, DEFIB SYNC is set to OFF as a default.
WARNING
z Improper defibrillation will endanger the patient’s safety. You user should decide
whether to defibrillate the patient based on the patient’s actual condition.
z Before defibrillating the patient, the user should ensure the defibrillator and the
monitor have been tested as a system and the two devices can work together safely
and effectively.
z Before defibrillating the patient, the user should ensure the defibrillator has been
connected to the auxiliary output port of the monitor, the defibrillation
synchronization function is enabled, and the filter mode is set to DIAGNOSTIC.
z When the defibrillation is done, sure to disable the defibrillation function and
select the filter method as needed.