C-13
C.2.14 System Alarm Messages
Alarm message A B L Cause Measure
REAL CLOCK
NEED SET
No No 1 The system time is incorrect. Reset the system time and then
restart the monitor.
REAL CLOCK
NOT EXIST
No No 1 No button battery, or the
battery power is depleted.
Add, or replace with a new
button battery.
KEYBOARD
INIT ERR N
No No 1
Note: N represents the error
number.
KEYBOARD
ERROR
No No 2
Keyboard error. The
keyboard cannot be used.
NET INIT ERR
(G.)
No No 2
NET INIT ERR
(Ram)
No No 2
NET INIT ERR
(Reg)
No No 2
NET ERR (Run
1)
No No 2
NET ERR (Run
2)
No No 2
The system cannot be
connected to the network
due to problems in the
monitor’s network part.
Contact our company for repair.
12V TOO HIGH No No 1
12V TOO LOW No No 1
A problem occurs to the
system power.
If this alarm message is given
for many times, contact our
company for repair.
BATTERY TOO
LOW
No No 1 The battery voltage is too
low.
Connect the monitor with AC
power to recharge the battery.