11-5
Figure 11-5 Neonate Sensor Placement (3)
NOTE
z If the sensor cannot be positioned accurately to the part to be measured, it may
result in inaccurate SpO2 reading, or the SpO2 even cannot be measured because
no pulse is detected. In this case, you must position the sensor again.
z The excessive patient movement may result in inaccurate reading. In this situation,
you must keep the patient quiet or change the measured position to reduce the
adverse influence of excessive movement.
WARNING
z In the process of extended and continuous monitoring, you should check the
peripheral circulation and the skin every 2 hours. If any unfavorable change takes
place, you should change the measured position in time.
z In the process of extended and continuous monitoring, you should periodically
check the position of the sensor. In case that the position of the sensor moves
during monitoring, the measurement accuracy may be affected.