A-18
> 8 Not specified
0 to 1
±0.15
1 to 5
±0.2
Enf, Iso, Hal
> 5 Not specified
0 to 25
±1
25 to 80
±2
O2 (Optional)
80 to 100
±3
2 to 60 BrPM
±1 BrPM
AwRR
> 60 BrPM Not specified
Drift meet the requirement of accurancy in 6 hours
Alarm range
CO2:
AwRR:
0 to 10 % (0 to 76 mmHg)
2 to 100 BrPM
Apnea alarm delay AwRR: 20 to 40 s
Refreshing rate 1s
Calibration Yearly calibration requested.
Calibration stability
After module being used for 12 consective months, the error is <
1%
CO2 250 ms (fall time 200 ms)
N2O 250 ms
O2 600ms
HAL, ISO, SEV, DES 300 ms
Rise time (10 % to 90 %)
Sampling flow 120ml/min,
using the DRYLINE™
water trap and neonatal
DRYLINE™ sampling
line (2.5m)
ENF 350 ms
CO2 250 ms (fall time 200 ms)
N2O 250 ms
O2 500ms
HAL, ISO, SEV, DES 300 ms
Rise time (10 % to 90 %)
Sampling flow 200ml/min,
using the DRYLINE™
water trap and adult
DRYLINE™ sampling
line (2.5m)
ENF 350 ms
Delay time < 4s
* Accuracy applies for the following conditions:
1. Measurements begin after the module warms up;
2. Ambient pressure is from 750 to 760 mmHg, and ambient temperature from 22 to 28ºC;
3. The measured gas is a dry gas and the balance gas N
2
;
4.Gas sample flow rate is 120 ml/min, respiration rate is 20 rpm with a fluctuation between ±3
rpm, and I:E is 1:2.
Vue de la page 243