12-3
5. Make sure that the cuff edge falls within the range of the <-> mark. If it does not, use a
larger or smaller cuff that will fit better.
6. The limb chosen for taking the measurement should be placed at the same level as the
patient's heart. If this is not possible, use the following method to correct the measurement
result:
If the cuff is placed higher than the heart level, add 0.75 mmHg (0.10 kPa) for each
centimeter of difference.
If it is placed lower than the heart level, deduct 0.75 mmHg (0.10 kPa) for each centimeter
of difference.
WARNING
z The width of the cuff should be either 40% of the limb circumference (50% for
neonates) or 2/3 of the upper arm length. The inflatable part of the cuff should
be long enough to circle 50-80% of the limb. The wrong size cuff can cause
erroneous readings. If the cuff size is in question, use a larger cuff.
z Do not apply the cuff to a limb that has an intravenous infusion or catheter in
place. This could cause tissue damage around the catheter when infusion is
slowed or blocked during cuff inflation.
z Make sure the air tubing connecting the blood pressure monitor is not blocked,
twisted, or tangled.
12.2.2 Operation Guides
1. To start a manual NIBP measurement
Access the NIBP SETUP menu and select MANUAL from the INTERVAL option; then,
press the NIBP key on the control panel to start a manual NIBP measurement; or
During the interval between two auto NIBP measurements, press the NIBP key on the
control panel to start a manual NIBP measurement.
2. To start auto NIBP measurement
Access the NIBP SETUP menu and select a time (e.g. 5MIN) from the INTERVAL options; press
the NIBP key on the control panel to start the auto NIBP measurment. When this measurment
finishes, the system will perform the NIBP measurment automatically as per the preset interval.
WARNING
z Auto non-invasive blood pressure measurements performed in long intervals
may incur ischemia and neuropathy in the limb wearing the cuff. When
monitoring a patient, examine the extremities of the limb frequently for normal
color, warmth and sensitivity. If any abnormality is observed, change the
position of the cuff on the patient or stop the blood pressure measurements
immediately.