C-10
Alarm message A B L Cause Measure
CO2 CALIB
ERROR
No No 2 Including the error
messages displayed at the
lower part of the menu
(like gas error,
measurement error,
zeroing failure, etc.)
Take measures specific to the
errors. For instance, in case
of a gas error, check if the gas
concentration is wrong. A
measurement error indicates
the module is removing the
occlusion or the filter line is
not connected properly.
CO2 INIT ERR Yes No 1 An error occurs during the
CO2 initialization.
CO2 COMM
STOP
No No 1 Failure in communication
between the CO2 module
and the main board.
CO2 COMM
ERR
Yes No 1 Error in communication
between the CO2 module
and the main board.
Restart the monitor. If the
error remains, contact our
company for repair.
C.2.12 AG Module Alarm Messages
Alarm message A B L Cause Measure
AG SENSOR OFF No Yes 3 The AG sensor is not
properly connected or
is disconnected.
Make sure the AG sensor
is properly connected.
AG NO WATERTRAP No Yes 3 The AG water trap is
disconnected from the
monitor.
Make sure the water trap
is properly connected.
CHANGE AG
WATERTRAP
Yes No 2 The AG water trap is
replaced.
Wait until the water trap
has been changed.
AG WATERTRAP TYPE
WRONG
Yes No 2 The AG water trap is
of a wrong type.
Replace with a water trap
of a correct type.
AG USA ERROR Yes No 2 AG module fault. Contact us to get rid of
the fault.
AG PARAMAGNETIC O2
ERR
Yes No 2
AG GALVANIC O2
ERROR
Yes No 1
AG OXIMA DEPLETION
WARN
Yes No 2
AG OXIMA DEPLETION
ERR
Yes No 2
O
2
module fault. Contact us to get rid of
the fault.
AG OCCLUSION Yes No 1 The actual pump rate
of the AG module is
<20ml/min, which
Remove the block of the
airway.