10-25
10.6.4 Arrhythimia Relearn
During ECG monitoring, you may need to start an arrhythmia relearn when a dramatic change in
the patient’s ECG pattern has occurred. A change in the ECG pattern could result in:
Incorrect arrhythmia alarms,
Loss of ST measurements, and/or
Inaccurate heart rate.
Arrhythmia relearn allows the monitor to learn the new ECG pattern to correct arrhythmia alarms
and heart rate value, and to restore ST measurements. To manually start an arrhythmia relearn,
select the ARR RELEARN option in the ARR ANALYSIS menu. The “ARR LEARNING”
message will be displayed in the information area.
10.6.5 Arrhythmia Alarm Setup
Select ARR ALARM >> in the ECG SETUP menu. The following menu appears.