Mindray DPM 1 Manuel de l'opérateur Page 149

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 274
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 148
11-7
11.2.5 SpO
2
Setup Menu
Selecting the SPO
2
label in the parameter window opens the following menu.
Figure 11-6 SpO2 Setup Menu
You can perform the following settings in this menu.
ALM SpO2 alarm on/off status
ON: When a SpO2 alarm occurs, the monitor gives alarm
indications and stores the alarm;
OFF: When a SpO2 alarm occurs, the monitor neither gives
alarm indications nor stores the alarm;
When OFF is selected, the
icon is displayed on the right
of the SPO2 label.
ALM LEV Alarm level
Options: HIGH and MED.
ALM REC Alarm recording
ON: When a SpO2 alarm occurs, the monitor enables the
recording;
OFF: When a SpO2 alarm occurs, the monitor does not
enable the recording.
SPO2 ALM HI SpO2 upper alarm limit
SPO2 ALM LO SpO2 lower alarm limit
PR ALM HI PR upper alarm limit
PR ALM LO PR lower alarm limit
Vue de la page 148
1 2 ... 144 145 146 147 148 149 150 151 152 153 154 ... 273 274

Commentaires sur ces manuels

Pas de commentaire