12-7
NS NM ND PR DATE TIME
108 84 70 68 2006-6-12 16:06:00
108 80 70 67 2006-6-12 16:11:00
108 80 70 72 2006-6-12 16:16:00
UNIT Options: mmHg, kPa;
INTERVAL Select MANUAL to set the monitor to manual NIBP
measurement mode, or select from the time options to
determine the interval between automatic measurements.
Optional intervals: 1, 2, 3, 4, 5, 10, 15, 30, 60, 90, 120, 180,
240, and 480MIN.
RESET Select this option to restore the initial settings of the
pressure pump. If the monitor fails to give a visual
indication when the pressure pump is working improperly,
selecting this option activates a self-test procedure, and
restores the monitor to normal performance.
CONTINUAL Select this option to start a continuous measurement for five
minutes.
CONFIG >> You can select this option to access the NIBP CONFIG
menu, in which you may select FACTORY CONFIG or
USER CONFIG. After finishing your selection and exiting
the menu, the system pops up a dialog box asking for your
confirmation.
12.4.1 Calibration
If you select the CALIBRATE option, the monitor starts the NIBP calibration and the
CALIBRATE option changes to STOP CAL. Selecting the option again stops the calibration.
Calibrate the cuff pressure reading with a calibrated reference manometer (or mercury
manometer) with accuracy higher than 1mmHg. To perform the calibration, follow the procedure
shown below:
1. Remove the blood pressure cuff from the monitor and replace it with a rigid metal container
or vessel with a capacity of 500 ml ± 5%.
2. Connect a calibrated reference manometer (with an error less than 1mmHg) and a ball pump
using “T” connectors as shown below.
3. Select the CALIBRATE option.
4. Inflate the metal container using the ball pump until the reference manometer reads 0, then
50, and finally 200 mmHg.
5. The difference between the indicated pressure of the reference manometer and the indicated
pressure of the monitor will not exceed 3 mmHg. Contact our Customer Service if these
values are not met.