18-3
18.2 SpO
2
Accessories
Extension Cable
Module type Part No.
Mindray DS SpO2 Module 0010-20-42594
SpO
2
Sensors
The SpO
2
sensor material that patients or other staff will come into contact with have undertaken
the bio-compatibility test and is verified to be in compliance with ISO 10993-1.
Mindray SpO2 Module
Type Model Patient Category Part No.
520A Adult (>30Kg) 520A-30-64101
520P Pediatric (10 to 50Kg) 520P-30-64201
520I Infant (3 to 20Kg) 520I-30-64301
Single
patient use
520N Neonate(<3Kg) 520N-30-64401
518B Adult, pediatric, neonate (multi-sites) 518B-30-72107
512E 512E-30-90390
512F
Adult (Finger type)
512F-30-28263
512G 512G-30-90607
Reusable
512H
Pediatric (Finger type)
512H-30-79061
The main cable must be used with the matching SpO2 sensor.
Wavelength emiited by the sensors intended for Mindray SpO
2
module: red light: 660 nm,
infrared light: 905 nm.
Wavelength emitted by the sensors intended for Nellcor SpO
2
module: red light: 660 nm,
infrared light: 890 nm.
The maximum photic output consumption of the sensor is less than 18 mW.
The information about the wavelength range and maximum photic output consumption can be
especially useful to clinicians, for example, clinicians performing photodynamic therapy.