A-1
A Product Specifications
A.1 Safety Classifications
Type of protection against
electric shock
Class I with internal electric power supply.
Where the integrity of the external protective earth (ground) in
the installation or its conductors is in doubt, the equipment shall
be operated from its internal electric power supply (batteries)
Degree of protection against
electric shock
CO2/AG:
ECG/RESP/TEMP/SpO2/NIBP/IBP:
BF (defibrillation proof)
CF (defibrillation proof)
Degree of protection against
hazards of ignition of
flammable anesthetic
mixtures
Not protected (ordinary)
Degree of protection against
harmful ingress of water
Not protected (ordinary)
Mode of operation Continuous
Equipment type Portable
A.2 Environmental Specifications
0 to 40
5 to 35 (With CO2 module)
Operating temperature
10 to 35 (With AION AG module)
Operating humidity 15 to 95%, noncondensing
Operating altitude
-500 to 4600 m (-1640 to 15092 feet)
-305 to 3014 m (-1000 to 9889 feet) (with CO2, AG or Nellcor
SpO2 module)
Storage temperature
-20 to 60
Storage humidity 10 to 95%, noncondensing
Storage and transportation
altitude
-500 to 13100 m (-1640 to 42979 feet)
-305 to 6096 m (-1000 to 20000 feet) (with CO2, AG or Nellcor
SpO2 module)