C-7
C.2.8 IBP Module Alarm Messages
Alarm message A B L Cause Measure
IBP1 SENSOR
OFF
No Yes 3 The invasive blood
pressure cable of channel 1
is disconnected from the
monitor.
Check if the IBP 1 sensor is
properly connected.
IBP2 SENSOR
OFF
No Yes 3 The invasive blood
pressure cable of channel 2
is disconnected from the
monitor.
Check if the IBP 2 sensor is
properly connected.
IBP1 NEED
ZERO-CAL
No No 3 The IBP transducer of
channel 1 has not been
zeroed.
Zero the IBP transducer of
channel 1.
IBP2 NEED
ZERO-CAL
No No 3 The IBP transducer of
channel 2 has not been
zeroed.
Zero the IBP transducer of
channel 2.
C.2.9 CO Module Alarm Messages
Alarm message A B L Cause Measure
TB SENSOR
OFF
No Yes 3 The cable for blood pressure
measurement is disconnected
from the monitor.
Check for proper
connection of TB cable.
C.2.10 Mindray DS CO
2
Module Alarm Messages
Alarm message A B L Cause Measure
CO2 NO
WATERTRAP
No No 2 The water trap is not
properly connected or is
disconnected.
Make sure the sidestream
CO
2
water trap is firmly
connected.
CO2 SENSOR
TEMP HIGH
No No 1 The temperature of the
sensor assembly is too
high.
CO2 SENSOR
TEMP LOW
No No 1 The temperature of the
sensor assembly is too
low.
CO2 AIRWAY
PRES TOO
HIGH
No No 2 The pressure inside the
airway is too high.
CO2 AIRWAY
PRES TOO
LOW
No No 2 The pressure inside the
airway is too low.
Restart the monitor if
necessary. If the problem
remains, contact our
company for repair.