Mindray DPM 1 Manuel de l'opérateur Page 217

  • Télécharger
  • Ajouter à mon manuel
  • Imprimer
  • Page
    / 274
  • Table des matières
  • MARQUE LIVRES
  • Noté. / 5. Basé sur avis des utilisateurs
Vue de la page 216
17-3
17.3 Disinfection and Sterilization
Sterilization or disinfection may cause damage to the equipment. We recommend the sterilization
and disinfection are contained in the hospital’s servicing schedule only when necessary. The
equipment should be cleaned prior to sterilization and disinfection.
Recommended sterilization material: Alcohol based (Ethanol 70%, Isopropanol 70%), and
aldehyde based.
WARNING
z Disinfection or sterilization may cause damage to the equipment; therefore,
when preparing to disinfect or sterilize the equipment, consult your hospital’s
infection controllers or professionals.
z The cleaning solutions above can only be used for general cleaning. If you use
them to control infections, we shall assume no responsible for the effectiveness.
NOTE
z ALWAYS dilute the solutions according to the manufacturer’s suggestions and
adopt lower concentration if possible.
z NEVER submerge the equipment into water or any solution, or pour water or
any solution on the equipment;
z ALWAYS wipe off all the excess liquids on the equipment surface and accessory
surface with a dry cloth;
z Never use EtO and formaldehyde to disinfect.
z Never permit high-pressure and high-temperature disinfection of the equipment
and accessories.
Vue de la page 216
1 2 ... 212 213 214 215 216 217 218 219 220 221 222 ... 273 274

Commentaires sur ces manuels

Pas de commentaire